Medical electrical equipment requires special precautions regarding electromagnetic compatibility
(EMC) and must be installed and operated according to the instructions in this manual. For more
section on page 54.
To maintain electromagnetic interference (EMI) within certified limits, the BladderScan
BVI 3000 system must be used with the cables, components, and accessories specified or
supplied by Verathon
®
. For additional information, see the
System Components & Accessories
sections. The use of accessories and/or cables other than those
specified or supplied may result in increased emissions and/or decreased immunity of the system.
The BladderScan BVI 3000 system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the system should be observed to verify normal
operation in the configuration in which it will be used.
This device can radiate radio frequency energy and is very unlikely to cause harmful interference
with other devices in the vicinity. There is no guarantee that interference will not occur in a particular
installation. Evidence of interference may include degradation of performance in this device or other
devices when operated simultaneously. If this occurs, try to correct the interference by using the
following measures:
• Turn devices on and off in the vicinity to determine the source of interference
• Reorient or relocate this device or other devices
• Increase the separation between devices
• Connect the device to an outlet on a circuit different than the other device(s)
• Eliminate or reduce EMI with technical solutions (such as shielding)
• Purchase medical devices that comply with IEC 60601‑1‑2 EMC Standards
Be aware that portable and mobile radio frequency communications equipment (cellular phones,
etc.) may affect medical electrical equipment; take appropriate precautions during operation.
CAUTION
WARNINGS
Do not use the BladderScan BVI 3000 instrument on:
• A patient who has open skin or wounds in the suprapubic area.
• A patient with ascites.
• A pregnant patient.
WARNING
Risk of explosion.
If you use the BladderScan BVI 3000 instrument in the presence of flammable
anesthetics, the hazard of potential explosion exists.