24
Chapter 2
January 2015
bk3000 User Guide
(16-01249-EN-01)
Non-Medical Equipment
Computer Security
When the 2300 Ultrasound System is connected to a hospital network, BK Medical
does not take any responsibility for computer viruses from the network that may
infect the bk3000.
Printers and
auxiliary
power outlet
ON/Standby
button does
not turn off
outlet
WARNING
ME-w2
An approved printer, specified in the Product Data sheet, can be connected to the internal
auxiliary power outlet on the system.
Do not use the auxiliary power outlet on the bk3000
ultrasound system for any other equipment.
The auxiliary power outlet is still live even when the ON/Standby button on the system is
turned off. To remove voltage from this outlet, you must unplug the power supply cord
from the power source.
WARNING
NME-w1
Follow the guidelines in EN 60601–1–1
(Safety requirements for medical electrical systems)
.
If you connect non-medical equipment (instruments that do not comply with safety
requirements for medical equipment, such as a video monitor, video recorder, endoscopic
camera control unit or other documentation device), this equipment must be placed
outside the patient environment (1.5m from the bed, for example). The equipment must
fulfill the relevant EN standard or other applicable national or international standard.
One of the following conditions must be fulfilled:
•
The system and other equipment are plugged into an external common isolation
transformer to control the leakage current during a ground connection fault.
or
•
The system is grounded with an additional safety ground connection (see “Additional
Protective Ground and Potential Equalization” on page 15).
If in doubt, contact your local BK Medical representative.
Check
external
media
Caution
CS-c1
You must perform a virus check on any external storage medium (USB device or DVD) to
make sure that it is virus-free before you connect it to the system.