bk3000 User Guide
(16-01249-EN-01)
Safety Information
21
Table 2-4. List of cables used in testing for EMC compliance
Do not attach transducers and other accessories unless the user guide for the
transducer or accessory states that it can be used with this system. Attaching other
equipment may cause an increase in electromagnetic emissions or may cause the
system to be more sensitive to electromagnetic interference.
Isolation of DICOM Network
The system must not be galvanically connected to a computer network (DICOM
®
)
that has not been isolated. If the network is not isolated, the system must be
connected via a network isolator DP0925 .
Wireless Networks
1
The system can be connected to a wireless network for printing and archiving data.
A Wi-Fi dongle and a 30 cm USB 3.0 extender cable are supplied with the system.
Connect the dongle to the extender cable and insert into one of the USB 3.0
connectors (see Fig 2-3 on page 18).
Connecting to a Wireless Network
You must establish a secure wireless network at your hospital, clinic or institution,
including a password for the network, before you can use the system’s Wi-Fi for
printing and archiving.
Connector name
Cable type
Type and length
DVI-I
Dual
link
Shielded,
5
m
S-video/composite Out
S-video
Shielded, 5 m
Audio In
Stereo, 3.5 mm jack
Shielded, 5 m
Audio Out
Stereo, 3.5 mm jack
Shielded, 5 m
USB
1
USB,
2.0
Shielded,
5
m
USB
2
USB,
2.0
Shielded,
5
m
USB
3
USB,
2.0
Shielded,
5
m
USB
4
USB,
2.0
Shielded,
5
m
USB 5
USB, 3.0
Shielded, 5 m
USB 6
USB, 3.0
Shielded, 5 m
10/100/1000 Ethernet
Network, CAT6E
Shielded, 5 m
1. Wi-Fi on the bk3000 has not been market cleared by the FDA or licensed by Health Canada.