63
4.
Select <Von Point> and press Enter.
Von Point
Von Latch
Exit
5.
Set the Von Point. Input 5 and press Enter to confirm.
Von Point Set
= 5V
6.
Press
to select <Von Latch> and press Enter. Set on/off state.
Select On and press Enter to confirm.
On
Off
7.
Press Esc key twice to exit menus.
8.
Press On/Off key to turn on the electronic load’s input. The
electronic load will begin sinking current at 5 V.
When in remote control mode, you can send SCPI command “VOLT:ON <n>” to set Von value; send
“VOLT:LATch ON” to enable Von LATCH function.
Save and Recall Operation
The stored settings of all channels can be recalled. The stored parameters include operation mode,
voltage/current values, slew rate, transient setting, and more. Up to 101 groups of setting
parameters can be saved. Group 0 can be used for power-up settings. Groups 1 through 100 can be
used for automatic testing parameters. All parameters are saved into non-volatile memory and will
remain saved even in a powered down state.
To Save
Panel Operations
VFD Display
1.
After setting up your parameters, press the Save key to save. It
will ask for the Save Group number.
Save Group
0
2.
Input the Save Group number (0-100) and press Enter to confirm.
Save Group
1