List Operation
List mode lets you generate complex sequences of input changes on a single channel with rapid,
precise timing, which may be synchronized with internal or external signals. This is useful when
running test sequences with a minimum amount of overhead.
The parameters of List operation include the name, number of steps (2-84), step width time (20us-
3600s), and every steps’ set value and slew rate. The list file can be saved in non-volatile memory
where it can be quickly recalled. Users can edit up to 7 groups of List files in CC mode only.
In List operation mode, the electronic load begins to enable the List operation when it receives the
trigger signal and will continue until the List operation is completed or the instrument receives
another trigger signal.
List Sequence
Figure 28 - List Mode Current Waveform
Setting Up List Operation Mode
The following is a quick tutorial of how to set up List Mode for your electronic load.
Panel Operations
VFD Display
1.
Power on the electronic load.
Self-test