Sound Level Calibration 

 

 

3. Features 

Conforms to ANSI S1.4-1984 and IEC 
60942-2003 Class 2. 
Calibration levels of 94dB and 114dB. 
Fits 1 inc h and 1/2 inch diameter microphone.   

4. Specifications 

Output sound pressure levels: 

94dB and 114dB re 20 uPa under reference 
conditions. 

Output frequency: 

/- 2%

 

Reference conditions: 

Temperature :23°C (74°F) 
Relative humidity: 50%  

Atmospheric pressure: 1013hpa 

Influence of ambient conditions: 

Temperature coefficient: 0.005dB/

?

 

Humidity coefficient: 0.005dB/%RH  

Total Harmonic Distortion (THD):

 

<3%

 

Accuracy of sound pressure level: 

+/- 0.5dB

 

Power:

 

one 9V battery 006P or IEC 6F22 or 

NEDA 1604. 

Battery life:

 

approx.40 hours (Alkaline Battery)

 

Battery test: 

Internal circuitry checks condition of battery 
continuously. 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for CAL73

Page 1: ...tput frequency 2 Reference conditions 2 Influence of ambient conditions 2 Total Harmonic Distortion THD 2 Accuracy of sound pressure level 2 Power 2 Battery life 2 Battery test 3 Dimensions 3 Weight 3 Ambient conditions 3 Storage temperature humidity 3 Accessories 3 5 Nomenclature And Function 4 6 Operating Preparation 5 7 Calibration Procedure 5 www com information itm com 1 800 561 8187 ...

Page 2: ...ed in this manual should only be performed by qualified personnel Periodically wipe the case with a dry cloth Do not use abrasives or solvents on this instrument Safety symbols Comply with EMC When servicing use only specified replacement parts 2 Description Sound level calibrator is used to calibrate sound level meters and other sound measurement equipment You can calibrate 1 inch diameter microp...

Page 3: ...ncy 1000Hz 2 Reference conditions Temperature 23 C 74 F Relative humidity 50 Atmospheric pressure 1013hpa Influence of ambient conditions Temperature coefficient 0 005dB Humidity coefficient 0 005dB RH Total Harmonic Distortion THD 3 Accuracy of sound pressure level 0 5dB Power one 9V battery 006P or IEC 6F22 or NEDA 1604 Battery life approx 40 hours Alkaline Battery Battery test Internal circuitr...

Page 4: ...e acceptable range Dimensions LxWxH 4 4 x 2 4 x 1 7inch 113 x 63 x 44mm Weight approx 170g inc luding battery Ambient conditions 0 40 C 32 140 F 10 90 RH 650 1080hpa Storage temperature humidity 10 50 C 14 122 F 0 70 RH Accessories Instruction manual carrying case 9V battery 1 2 microphone adapter www com information itm com 1 800 561 8187 ...

Page 5: ...cate LED Power and output level select switch ƒ Transducer assembly 1 inch cavity for microphone insertion 1 2 inch microphone adapter Battery cover 94dB 114dB ANSI S1 40 1984 IEC 60942 Class 2 OFF NEDA 1604 9V BAT TERY 6F22 006P 1 2 3 4 5 www com information itm com 1 800 561 8187 ...

Page 6: ...20 You can calibrate in noisy environments and check linearity 7 Calibration Procedure 1 The cavity of the calibrator will accommodate 1 inch microphone 2 When the calibration is performed to an instrument with 1 2 inch microphone the 1 2 inch microphone adaptor will have to be inserted by gently pushing it into the cavity till the end 3 Place the sound level calibrator over the microphone of the ...

Page 7: ...rument may be set to FAST or SLOW response and C or A weighting 5 Read the level on the instrument under test and adjust the sensitivity control for the correct indication of the sound level calibrator level selected in step 6 When the calibrator is not being used please switch OFF the power to save the battery CAUTION Ambient sources of noise or vibration can cause a false calibration indication ...

Reviews: