Sound Level Calibration 

 

 

being calibrated. 

(4) On the instrument under test, set the level range 

control to the range having 100dB as its upper 

limit if 94dB was selected on the sound level 
calibrator. If 114dB was selected on the sound 
level calibrator, choose a range with on upper 
limit of 120dB. The instrument may be set to 

FAST or SLOW response and C or A weighting. 

(5) Read the level on the instrument under test and 

adjust the sensitivity control for the correct 

indication of the sound level calibrator level 
selected in step. 

(6) When the calibrator is not being used, please 

switch OFF the power to save the battery. 

 
 

CAUTION! 

Ambient sources of noise or vibration can 
cause a false calibration indication; this can be 
especially signification at the lower 84dB level. 

 
 
 
 
 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for CAL73

Page 1: ...tput frequency 2 Reference conditions 2 Influence of ambient conditions 2 Total Harmonic Distortion THD 2 Accuracy of sound pressure level 2 Power 2 Battery life 2 Battery test 3 Dimensions 3 Weight 3 Ambient conditions 3 Storage temperature humidity 3 Accessories 3 5 Nomenclature And Function 4 6 Operating Preparation 5 7 Calibration Procedure 5 www com information itm com 1 800 561 8187 ...

Page 2: ...ed in this manual should only be performed by qualified personnel Periodically wipe the case with a dry cloth Do not use abrasives or solvents on this instrument Safety symbols Comply with EMC When servicing use only specified replacement parts 2 Description Sound level calibrator is used to calibrate sound level meters and other sound measurement equipment You can calibrate 1 inch diameter microp...

Page 3: ...ncy 1000Hz 2 Reference conditions Temperature 23 C 74 F Relative humidity 50 Atmospheric pressure 1013hpa Influence of ambient conditions Temperature coefficient 0 005dB Humidity coefficient 0 005dB RH Total Harmonic Distortion THD 3 Accuracy of sound pressure level 0 5dB Power one 9V battery 006P or IEC 6F22 or NEDA 1604 Battery life approx 40 hours Alkaline Battery Battery test Internal circuitr...

Page 4: ...e acceptable range Dimensions LxWxH 4 4 x 2 4 x 1 7inch 113 x 63 x 44mm Weight approx 170g inc luding battery Ambient conditions 0 40 C 32 140 F 10 90 RH 650 1080hpa Storage temperature humidity 10 50 C 14 122 F 0 70 RH Accessories Instruction manual carrying case 9V battery 1 2 microphone adapter www com information itm com 1 800 561 8187 ...

Page 5: ...cate LED Power and output level select switch ƒ Transducer assembly 1 inch cavity for microphone insertion 1 2 inch microphone adapter Battery cover 94dB 114dB ANSI S1 40 1984 IEC 60942 Class 2 OFF NEDA 1604 9V BAT TERY 6F22 006P 1 2 3 4 5 www com information itm com 1 800 561 8187 ...

Page 6: ...20 You can calibrate in noisy environments and check linearity 7 Calibration Procedure 1 The cavity of the calibrator will accommodate 1 inch microphone 2 When the calibration is performed to an instrument with 1 2 inch microphone the 1 2 inch microphone adaptor will have to be inserted by gently pushing it into the cavity till the end 3 Place the sound level calibrator over the microphone of the ...

Page 7: ...rument may be set to FAST or SLOW response and C or A weighting 5 Read the level on the instrument under test and adjust the sensitivity control for the correct indication of the sound level calibrator level selected in step 6 When the calibrator is not being used please switch OFF the power to save the battery CAUTION Ambient sources of noise or vibration can cause a false calibration indication ...

Reviews: