Common Command Introduction
21
Query frequency counter information:
FCNT?
Return:
FCNT STATE,ON,FRQ,10000000HZ,DUTY,59.8568,REFQ,
1e+07HZ,TRG,0V,PW,5.98568e-08S,NW,4.01432e-08S,FR
QDEV,0ppm,MODE,AC,HFR,OFF
2.27 Invert Command
Description
Sets or gets polarity of current channel.
Command Syntax
<channel>:INVerT <parameter> <channel>={C1, C2} <parameter>={ON, OFF}
Query Format
<channel>:INVerT?
<channel>={C1, C2}
Query Response
<channel>:INVerT <parameter>
Example
<channel>:INVerT <parameter>
Set C1 ON:
C1:INVT ON
Read the polarity of channel one.
C1:INVT? Return:
C1:INVT ON
2.28 Coupling Command
Description
Sets or gets channel coupling parameters. You can only set coupling value when trace switch off.
Command Syntax
COUP (COUPling)<parameter>
<parameter>={a parameter from the table below}
Value Parameters Description
TRACE <trace>
Trace switch
STATE <state>
State of channel coupling.
BSCH <bsch>
Value of base channel.
FDEV <frq_dev> Value of f frequency deviation.
PDEV <pha_dev> Value of position phase deviation.
FCOUP <fcoup>
Value of frequency coupling switch
FRAT <frat>
Value of frequency coupling ratio
PCOUP <pcoup>
Value of phase coupling switch
PRAT <prat>
Value of phase coupling ratio
ACOUP <acoup>
Value of amplitude coupling switch
ARAT <arat>
Value of amplitude coupling ratio
ADEV <adev>
Value of amplitude coupling deviation
where: <trace>={ON, OFF}
< state >={ON, OFF}
< bsch >={CH1, CH2}
< frq_dev >={Default unit is “Hz”, value range depends on the model}
< pha_dev >={Default unit is “
°
”value range depends on the model}
<fcoup>,<acoup>,<pcoup>={ON, OFF}
<frat>,<prat>,< arat >={a ratio value. value range depends on the model}
<adev>={a deviation value. value range depends on the model}
Query Format
COUP?