5
F E A T U R E S
T R I G G E R I N G F E A T U R E S
Two Trigger Modes
Selectable normal (triggered) or automatic sweep
modes.
Triggered Sweep
Sweep remains at rest unless adequate trigger signal
is applied. Fully adjustable trigger level and (+) or ()
slope.
AUTO Sweep
Selectable AUTO sweep provides sweep without
trigger input, automatically reverts to triggered
sweep operation when adequate trigger is applied.
Five Trigger Sources
Five trigger source selections, including CH 1, CH 2,
alternate, EXT, and LINE.
Video Sync
Frame (TV V) or Line (TV H) triggering selectable
for observing composite video waveforms. TV-H
position can also be used as low frequency reject
and TV-V position can be used as high frequency
reject.
Variable Holdoff
Trigger inhibit period after end of sweep adjustable. Permits
stable observation of complex pulse trains.
O T H E R F E A T U R E S
X Y Operation
Channel 1 can be applied as horizontal deflection (X-axis)
while channel 2 provides vertical deflection (Y-axis).
Built-in Probe Adjust Square Wave
A 0.5 V p-p, 1 kHz square wave generator permits probe
compensation adjustment.
Channel 2 (Y) Output
A buffered 50
Ω
output of the channel 2 signal is available at
the rear panel for driving a frequency counter or other
instruments. The output is 50 mV/div (nominal) into 50
Ω
.
Supplied With Two Probes