background image

 

 

23

APPENDIX II 

UNIQUE CHARACTERISTICS OF DIGITAL STORAGE OSCILLSCOPES 

 

Digital Storage Oscilloscopes (DSO’s) use a digital sampling 

technique to convert analog input signals to a series of digital 
words that can be stored in memory. Since digital sampling has 
disadvantages as well as advantages, it is important to be aware 
of these unique characteristics of DSO’s. 

ALIASING 

This DSO uses Real Time Sampling when sweep TIME/DIV 

settings of 50 sec/div to 20 µs/div are selected. Real Time 
Sampling simply means that samples of the input signals are 
taken at equal spaces (e.g., every 0.25 ms when the 50 ms/div 
range is selected). With Real Time Sampling, a phenomena 
called “Aliasing” can occur when the input signal is not 
sampled often enough. This causes the digitized signal to appear 
to be of a lower frequency than that of the input signal. Unless 
you have an idea what the input signal is supposed to look like, 
you will usually be unaware that Aliasing is occurring. 

A l i a s i n g   E x a m p l e  

To see what actually occurs when a Digital Storage Oscil-

loscope is Aliasing, perform the following example. 

1.

 

Apply a 10 kHz signal to the input jack and set the sweep 
TIME/DIV control to 50 µs/div. You should see about 5 
cycles of the waveform on the display. Since the DSO 
samples the input waveform 200 times per division, each 
cycle is sampled 400 times. 

2.

 

Now change the sweep TIME/DIV control to 2 ms/div. 
The display should look crowded. Because the DSO takes 
200 samples per division, the sample points are 10 µs 
apart. Since the input signal is at a frequency of 10 kHz, it 
is being sampled 10 times per cycle. The resulting display 
is too crowded to be useful, however, it is not incorrect (it 
is very similar to what you would see on a conventional 
analog oscilloscope). 

3.

 

Change the setting of the sweep TIME/DIV control to 20 
ms/div. Vary the frequency a slight amount (until the 
display is readable) to obtain as few cycles as possible on 
the CRT. If you were to calculate the frequency of the 
signal from the display, you would come up with a much 
lower frequency than that of the actual frequency of the 
signal at the input jack. As an example, if three cycles are 
displayed, the calculated frequency would be 
approximately 15 Hz. This is 

obviously incorrect. This occurs because the DSO is 
taking one sample every 0.1 ms and a 10 kHz signal has 
one cycle every 0.1 ms. What is actually happening is that 
the frequency is off (not perfectly 10 kHz) by just enough 
to cause the DSO to take one sample at a slightly 
different place on each cycle of the waveform. 

A v o i d i n g   A l i a s i n g  

Aliasing is not limited to the above example. This phe-

nomenon can occur anytime that at least two samples per cycle 
are not taken (whenever the sweep TIME/DIV setting is much 
too slow for the waveform being applied to the input). 
Whenever the frequency of the signal is unknown, always 
begin with the fastest real-time sweep speed (20 µs/div) or by 
viewing the waveform in the analog mode of operation first. 

N O T E  

Viewing one-time events or glitches is not 
possible when sweep TIME/DIV settings 
higher than 20 s/div are selected. Viewing 
one-time events poses no problem with Ali-
asing because Aliasing can occur only with 
repetitive waveforms. 

EQUIVALENT TIME SAMPLING 

This oscilloscope uses a sampling method called Equivalent 

Time Sampling when sweep speeds higher than 20 µs/div are 
selected. This method permits viewing of repetitive waveforms 
to 20 MHz, although the maximum digital sampling rate is 10 
Msamples/sec. When the Equivalent Time Sampling mode is 
activated, one sample is taken during each cycle. Of course if 
one sample is taken during each cycle at the trigger point (the 
same point on each cycle), only a flat trace would be produced. 
Therefore, it is necessary to take each sample further (in time) 
from the trigger point than the last sample. This incremental 
delay is deter-mined by the sweep TIME/DIV control setting. 
Because 1024 (1 k) samples are needed to fill the display, the 
oscilloscope must sample 1024 cycles of the waveform. 

Therefore, only repetitive waveforms should be observed in 

this mode. Irregularities that are present on an otherwise 
repetitive waveform are not likely to show up when the 
Equivalent Time Sampling method is used. With only one 
sample being taken during each cycle, it is very likely that 
glitches and other irregularities will be skipped over. 

 
 
 
 
 
 

Summary of Contents for 2522C

Page 1: ...20 MHz DIGITAL STORAGE ANALOG OSCILLOSCOPE INSTRUCTION MANUAL Model 2522C...

Page 2: ...wet 4 Use the time proven one hand in the pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return...

Page 3: ...AL STARTING PROCEDURE 14 SINGLE TRACE DISPLAY 14 DUAL TRACE DISPLAY 14 TRIGGERING 15 MAGNIFIED SWEEP OPERATION 17 X Y OPERATION 17 VIDEO SIGNAL OBSERVATION 17 DIGITAL STORAGE OPERATION 17 MAINTENANCE...

Page 4: ...r CRT with large 8 x 10 centimeter viewing area Convenience Trace rotation electrically adjustable from front panel 0 10 90 and 100 markers for rise time measurements DUAL TRACE FEATURES Dual Trace Mo...

Page 5: ...for observing composite video waveforms TV H position can also be used as low frequency reject and TV V position can be used as high frequency reject Variable Holdoff Trigger inhibit period after end...

Page 6: ...oximately 500 kHz Polarity Reversal CH 2 invert Maximum Input Voltage 400 V dc ac peak HORIZONTAL AMPLIFIER Input through channel 1 input X Y mode CH 1 X axis CH2 Y axis Sensitivity Same as vertical c...

Page 7: ...ce Equivalent Time Sampling Bandwidth 20 MHz for repetitive waveforms Dot Joining Linear interpolation between samples Digital Display Modes Roll Stored data and display updated continually Refresh St...

Page 8: ...er cord summed and displayed as a single signal If the Channel 2 POSition PULL INVert control is pulled out the input from channel 2 is subtracted 7 CAL Terminal Terminal provides 0 5 V p p 1 kHz nomi...

Page 9: ...tion Adjusts vertical position of channel 1 trace PULL CHOP When this switch is pulled out in the dual trace mode the channel 1 and channel 2 sweeps are chopped and displayed simultaneously normally u...

Page 10: ...less of the VERTical MODE switch setting ALTernate Selects alternate triggering used in dual trace 19 Time Base TIME DIV Control Provides step selec mode permits each waveform viewed to become its 20...

Page 11: ...N inum normal The holdoff period increases pro gressively with clockwise rotation TRIGger LEVEL Trigger level adjustment determines the point on the triggering waveform where the sweep is trig gered R...

Page 12: ...and SLOW X100 if engaged ROLL mode is applicable only at time base settings of 10 ms div to 0 5 S div 33 SINGLE Pushbutton Engaging this switch enables the single sweep mode of digital storage operati...

Page 13: ...o waveform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never allow a small spot of high brilliance to remain stationary on the...

Page 14: ...red POWER pushbutton 3 A trace should appear on the CRT Adjust the trace brightness with the INTENSITY control and the trace sharpness with the FOCUS control SINGLE TRACE DISPLAY Either channel 1 or c...

Page 15: ...y when first setting up the scope to observe a waveform it provides sweep for wave form observation until other controls can be properly set Once the controls are set operation is often switched back...

Page 16: ...SLOPE control is set to the out position a sweep trigger is developed from the trigger source waveform as it crosses the threshold level in a negative going direction TIME BASE Control Set the Time B...

Page 17: ...g fields of video and 5 ms div for complete frames two interlaced fields of video At most points of measurement a composite video signal is of the polarity that is the sync pulses are negative and the...

Page 18: ...ator goes out Single sweep mode can be used in the presence of very rapidly occurring events even continuous waveforms if desired In that case the period between pushing the RE SET switch and the arri...

Page 19: ...es 1000 ms 1 sec div when this switch is engaged This time base expansion is extremely useful for observing very slow events With it the scope is capable of recording an event up to 500 seconds in dur...

Page 20: ...e Pry fuseholder away from receptacle 3 When reinstalling fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION T...

Page 21: ...skills and test equipment required for instrument repair and calibration many customers prefer to rely upon B K Precision for this service To use this service even if the oscilloscope is no longer un...

Page 22: ...sured at the fastest sweep speed and using X10 magnification This sweep rate is 10 ns div A rise time of less than about five divisions at this sweep speed should be calculated Calculated Measurements...

Page 23: ...equency of the signal at the input jack As an example if three cycles are displayed the calculated frequency would be approximately 15 Hz This is obviously incorrect This occurs because the DSO is tak...

Page 24: ...Intentionally left blank...

Page 25: ...7 www bkprecision com 714 921 9095 Include with the returned instrument your complete return shipping address contact name phone number and description of problem LIMITED ONE YEAR WARRANTY B K Precisi...

Page 26: ...22820 Savi Ranch Parkway Yorba Linda CA 92887 www bkprecision com 2009 B K Precision Corp 481 315 9 001 Printed in Taiwan...

Reviews: