BK Medical 20R3 User Manual Download Page 1

 
16-01278-EN-07
May 2019

For Professional Users Only

20R3 Transducer 

User Guide

Summary of Contents for 20R3

Page 1: ...16 01278 EN 07 May 2019 For Professional Users Only 20R3 Transducer User Guide...

Page 2: ...dical transducer contains information about the date of manufacture BK Medical Customer Satisfaction Input from our customers helps us improve our products and services Your opinions are important to...

Page 3: ...sing a Transducer Cover 9 Selecting the Crystal and Imaging Frequency 10 Anatomical Position Insertion Depth and Crystal Position 10 Image Orientation 11 Changing Orientation 11 Endorectal Imaging 11...

Page 4: ...4...

Page 5: ...information Intended use The transducer is intended for diagnostic ultrasound imaging or fluid flow analysis of the human body Indications for use 20R3 is suitable for transrectal and transvaginal im...

Page 6: ...ted after the transducer number at the top of the screen H high frequency or L low frequency The crystal assembly can be positioned manually or if you have a 3D option installed see 3D Imaging on page...

Page 7: ...ectromagnetic compatibility for this transducer can be found in the Product Data Sheet and the Technical Data BZ2100 that accompany this user guide Forward In button Backward Out button Start Stop Cap...

Page 8: ...repair of BK electromedical equipment must be carried out only by the manufacturer or its authorized representatives BK Medical reserves the right to disclaim all responsibility including but not limi...

Page 9: ...merica it is recommended to use transducer covers that have been market cleared In Canada use only licensed transducer covers In Europe transducer covers must be CE marked Apply sterile gel to the tip...

Page 10: ...plane is in the patient by setting anatomical position insertion depth and crystal position To set the insertion depth 1 Find the actual insertion depth by reading the scale on the transducer Point at...

Page 11: ...uctions Endorectal Imaging Water Standoff Systems You must use a water standoff system to image the rectum from the anorectal junction and beyond BK Medical offers reusable water standoff systems and...

Page 12: ...nents NOTE See the Product Data sheet for ordering information Preparing the Patient 1 Always perform a digital palpation before you insert a rectosigmoidoscope or endo transducer into the rectum 2 Gi...

Page 13: ...rker on the transducer 4 Pull a water standoff cover over the transducer attaching it to the groove on the water standoff collar Check the water standoff cover for tears 5 Secure the water standoff co...

Page 14: ...ssible 9 Refill the water standoff cover with degassed water Repeat this procedure until there is no air left in the water standoff cover 10 Remove enough water from the water standoff cover to allow...

Page 15: ...position 2 Apply water to moisten the transducer Figure 7 Moistening the transducer with water 3 Pull the inner cover A in Fig 5 over the transducer See Fig 8 Check the inner cover for damage tears Ca...

Page 16: ...cm together The tape will be too short to fit around the transducer 5 Push the non handle end of the tape through the opening in the inner cover and wind the tape tightly around the cover and then aro...

Page 17: ...cover B in Fig 5 over the transducer and inner cover Make sure that it completely covers the tape as shown in Fig 12 Figure 12 Pulling the outer cover on to cover the tape completely 7 Using a standar...

Page 18: ...rface of the standoff that will enter the patient Aspirate enough water from the standoff to allow for easy insertion The transducer is now prepared for an endorectal examination Performing a Rectosco...

Page 19: ...t it just covers the area you want to image WARNING Colo w1 Donotuseexcessiveforceduringinsertion Donotmakeexcessivelateralmovementsduring or after insertion Risk of injury or tissue damage to the pat...

Page 20: ...e transducer if you are using either the UA0672 or UA0673 rectosigmoidoscope If you are using the single use rectosigmoidoscope UA0683 then it will be at the 190 mm mark Hold the transducer in positio...

Page 21: ...page 23 If you have not imaged the entire object of interest in the first pass pull the transducer back another 4 6 cm and image again Removing the Reusable Water Standoff System from the Patient Reus...

Page 22: ...gure 16 Removing the inner cover Clean Immediately after Use WARNING Reproc w3 Immediately after use you mustpre cleanthe device until visually clean including device lumens if existing Conduct the th...

Page 23: ...e system user guide for instructions NOTE The maximum length of a dataset acquired using this transducer is 60mm The resolution of the dataset is never worse than 0 5mm i e 2 images per 1mm of movemen...

Page 24: ...24 May 2019 20R3 User Guide 16 01278 EN 07...

Page 25: ......

Page 26: ......