BT-720 Operation manual
P/N:720-ENG-OPM-ENG-R03
Bistos Co., Ltd.
2020.10
5
Indicates to keep the device dry.
Indicates the medical device that can be broken or damaged if not handled
carefully.
Indicates to keep upright
Indicates the maximum stacking limit.
Indicates the temperature limitation for operation, transport and storage.
Indicates the humidity limitation for operation, transport and storage.
Indicates the range of atmospheric pressure to which the medical device
can be safely exposed.
Indicates the device contains natural rubber latex.(Accessories)
Indicates the packing material is recyclable.
Indicates to not dispose the device together with unsorted municipal
waste(for EU only). The solid bar symbol indicates that mains adapter is
put on the market after 13 August 2005.
0.1 General precautions, warnings and cautions
•
Examine the patient monitor and any accessories periodically to ensure that the cables,
adapter cords and instruments do not have visible evidence of damage that may affect
patient safety or performance. The recommended inspection interval is once per week or
less. Do not use the patient monitor if there is any visible sign of damage.
•
Only the DC power adapter supplied with the BT-720 is approved for use with the device.
•
Do not attempt to service the BT-720 patient monitor. Only qualified service personnel by
Bistos Co. Ltd. should attempt any needed internal servicing.
•
Perform periodic safety testing to insure proper patient safety. This should include
leakage current measurement and insulation testing. The recommended testing interval is
once per year.
•
If the hospital or healthcare institutions using this device fail to implement a satisfactory
maintenance schedule, it will result in device failure and may endanger the patient’s
safety.
•
Use the patient monitor under the conditions specified in this operation manual. Beyond
the conditions, the patient monitor may not function properly and the measurement
results may not accurate and may result in device failure or endangering the patient’s
safety.
•
Do not operate the BT-720 patient monitor if it fails to pass the power on self-test
procedure.
•
During the operation, do not disconnect any cable.
•
The BT-720 patient monitor is intended to be used by clinical professionals or trained
doctors, nurses or laboratory assistant.
•
Do not service and maintain or clean the device including accessories while in use with a
patient.
•
Using the device to one patient at a time.