BT-720 Operation manual
P/N:720-ENG-OPM-ENG-R03
Bistos Co., Ltd.
2020.10
13
6
Perfusion Index. Display the measured perfusion index.
7
Sweep speed. Display the user set SpO
2
waveform speed.
8
RTIP. Display the current measured cuff pressure value.
9
MAP. Display the average pressure value.
10
Diastolic blood pressure value.
11
Systolic blood pressure value.
12
The unit of NIBP.
13
Battery status.
14
Patient type.
15
Display current time.
16
The unit of SpO
2
.
1.7 Essential performance
This device Patient Monitor provides various patient vital signs such as pulse rate, blood oxygen
saturation and non-invasive blood pressure by placing the sensors to the appropriate site of patient.
The device is composed with display, control circuit and panel, and input part for various sensors. It
detects SpO2, PR and NIBP using specific sensors and cuff. The detected analog signal amplifies and
converted to digital. This concerted data feed to the CPU and converted to the display format as
number and waveform. This device is incorporated with alarm system. The alarm generated when the
detected signal range is beyond the user set alarm limits.
2 Preparing for operations
2.1 Installation
To ensure normal working of the monitor, read this chapter before use, and install as required.
WARNING
All analog and digital devices connected to the monitor must be certified by IEC
standards (e.g. IEC 60950 Data processing equipment standard and IEC 60601-1 Medical
equipment standard). Furthermore, all configurations shall comply with valid version of
IEC 60601-1 standard. The personnel connecting additional devices to the input / output
signal ports are responsible for the compliance with IEC 60601-1 standard. If there is any
question, please contact Bistos.
If the patient cable interface and network interface are connected with multiple devices,
the total electric leakage current cannot exceed the allowable value.
The copyright of monitor software belongs to Bistos. Without permission, any
organization or individual shall not interpolate, copy or exchange by any means or form.
When the monitor is combined with other devices, it must comply with IEC 60601-
1:2005 + A1:2012, and should not be connected with multi-socket outlet or extension
cord.
Do not connect the device on other equipment or network, to which a signal
input/output part may be connected.
Prior to installation, the operator must ensure that the following space, power, environmental
requirements are met.
2.1.1 Unpack and check
BT-720 patient monitor was inspected rigorously at the factory before delivery, in order to avoid being
hit when transported, carried out careful packaging. Before unpacking, carefully inspect the package. If
any damage, please immediately contact the Bistos. Unpack in the correct way, carefully remove the
monitor and accessories from the box and check with the packing list. Check if there is any mechanical
damage, the all listed are completely packed. If you have questions, please contact the marketing
department of Bistos or agency.
Please keep the packing box and materials for use in future transporting or storage.