Evia Technical Manual 43
Table 9: Percentages of Patients Meeting Success Criteria
Criteria
% Patients
Slope > 0.825
53.8% (28/52)
Slope > 0.650
78.8% (41/52)
Within 20 bpm of MCLR
80.8% (42/52)
Achieved MCLR
46.2% (24/52)
Figure 5
shows the obtained heart rate versus the expected
heart rate during the CAEP treadmill test for all patients
completing at least 4 stages of exercise (n = 52).
CAEP Workload (normalized)
1.0
.8
.6
.4
.2
0.0
O
bt
ained Heart
Rat
e (norm
alized)
1.4
1.2
1.0
.8
.6
.4
.2
0.0
Upper 95% CI
Lower 95% CI
Mean (n=52)
Expected Wilkoff HR
Figure 5. Obtained vs. Expected Heart Rate during CAEP