![BIOTRONIK Evia DR Technical Manual Download Page 119](http://html1.mh-extra.com/html/biotronik/evia-dr/evia-dr_technical-manual_2756148119.webp)
Evia Technical Manual 111
8.2 Timing Statistics
8.2.1 Event Counter
8.2.2 Event Counter
The event counter totals all of the sensed and paced events.
With the event counter, the following events and event
sequences can be registered over several years:
•
Atrial:
o
As, atrial sensed events
o
As (PVARP), atrial events sensed in the PVARP
window
o
Ars, atrial events sensed in the ARP
o
As (FFP), atrial events sensed in the far field
protection period
o
Ap, atrial paced events
•
Ventricular:
o
Vs, ventricular sensed events
o
PVC, ventricular sensed event not preceded by
an atrial sensed event.
o
Vrs, ventricular sensed events that occur within
the ventricular refractory period.
o
Vp, ventricular paced events
N
OTE
:
All event counter data are transmitted to the programmer
and evaluated there, but not all events are displayed in detail
on the programmer.