Evia Technical Manual 105
7.15.4 Types of Report Transmissions
When the Home Monitoring function is activated, the
transmission of a report (Cardio Report) from the implant can be
triggered as follows:
•
Trend report – the time period (daily) initiates the report
•
Event report – the pulse generator detects certain
events, which initiate a report
7.15.4.1 Trend Report
The time of the report transmission is programmable. For
periodic messages, the time can be set anywhere between 0:00
and 23:59 hours. It is recommended to select a time between
0:00 and 4:00.
The length of the time interval (monitoring interval) is preset to
“daily”. For each monitoring interval, a data set is generated in
the implant and the transmission is initiated at the designated
time.
7.15.4.2 Event Report
When certain cardiac and technical events are detected by the
implant, a report transmission is automatically triggered. This is
described as an “event message”.
The following clinical and technical events initiate a Home
Monitoring message transmission:
•
Atrial Lead Check < 100+/-50 Ohm and > 2500+/-500
Ohm
•
Ventricular Lead Check < 100+/-50 Ohm and > 2500+/-
500
•
VCC
Disabled
•
ERI
detected
•
High
Ventricular rate
•
Atrial tachyarrhythmia persisting beyond a
programmable time limit or
Summary of Contents for Evia DR
Page 1: ...Evia Family of Implantable Pulse Generators Technical Manual...
Page 8: ...vi Evia Technical Manual...
Page 13: ...Evia Technical Manual 5...
Page 15: ...Evia Technical Manual 7...
Page 28: ...20 Evia Technical Manual...
Page 128: ...120 Evia Technical Manual...
Page 137: ...Evia Technical Manual 129...
Page 143: ...Evia Technical Manual 135...
Page 152: ...144 Evia Technical Manual...
Page 156: ...148 Evia Technical Manual...
Page 160: ...152 Evia Technical Manual...
Page 182: ......