168 Evia Technical Manual
DVI
The DVI mode is
clearly indicated
if
•
AV sequential contraction is necessary due to
symptomatic bradycardia and slow atrial rate, and/or
•
a pacemaker syndrome has already been documented.
The DVI mode is conditionally indicated for
•
frequent supraventricular arrhythmia in which a
combination of pacing and medication has proved
therapeutically effective, and/or
•
the presence of a bradycardia-tachycardia syndrome,
presuming that setting the atrial rate and the AV interval
with or without accompanying medication stops or
prevents supraventricular arrhythmia.
The DVI mode is
contraindicated
for
•
frequent or persistent supraventricular tachyarrhythmia,
including atrial fibrillation or flutter.
Dual Chamber Modes
VDD
The VDD mode is
clearly indicated
for
•
ventricular pacing when adequate atrial rates and
adequate intracavitary complexes are present. The
indication includes the presence of complete AV block
when
1. the atrial contribution is necessary for hemodynamic
optimization, and/or
2. a pacemaker syndrome has already occurred or is
expected.
Summary of Contents for Evia DR
Page 1: ...Evia Family of Implantable Pulse Generators Technical Manual...
Page 8: ...vi Evia Technical Manual...
Page 13: ...Evia Technical Manual 5...
Page 15: ...Evia Technical Manual 7...
Page 28: ...20 Evia Technical Manual...
Page 128: ...120 Evia Technical Manual...
Page 137: ...Evia Technical Manual 129...
Page 143: ...Evia Technical Manual 135...
Page 152: ...144 Evia Technical Manual...
Page 156: ...148 Evia Technical Manual...
Page 160: ...152 Evia Technical Manual...
Page 182: ......