![BIOTRONIK Enitra 8 HF-T Technical Manual Download Page 32](http://html1.mh-extra.com/html/biotronik/enitra-8-hf-t/enitra-8-hf-t_technical-manual_2756130032.webp)
32
Parameters
Rate Adaptation
Rate Adaptation
CLS modes: closed loop
stimulation
Parameters valid for 8 series devices:
R modes: Accelerometer
Parameters valid for devices with R modes:
Parameter
Range of values
Standard
SR
DR HF
Maximum CLS rate
80 ... (10) ... 160 bpm
120 bpm
x
x
x
CLS response
Very low; Low; Medium;
High; Very high
Medium
x
x
x
CLS resting rate
control
OFF; +10 ... (10) ... +50 bpm
+20 bpm
x
x
x
Vp required
Yes; No
No
When BiV is
set: Yes
x
x
x
Parameter
Range of values
Standard
SR
DR HF
Sensor gain
AUTO; Very low; Low;
Medium; High; Very high
AUTO
x
x
x
Max. activity rate
80 ... (10) ... 180 bpm
120 bpm
x
x
x
Sensor threshold
Very low; Low; Medium;
High; Very high
Medium
x
x
x
Rate fading
OFF; ON
OFF
x
x
x
Rate increase
1; 2; 4; 8 bpm/cycle
2 bpm/cycle
x
x
x
Rate decrease
0.1; 0.2; 0.5; 1.0 bpm/cycle
0.5
bpm/cycle
x
x
x