Implantation
Replacement Indications
Magnet response at ERI
After reaching ERI, pacing is performed as follows after applying the magnet or
programming head:
Expected service times
after ERI
The information is based on the following:
•
Lead impedance of 500 Ω or 600 Ω
•
100% pacing
•
Interval from ERI to EOS for the single-chamber device in AAI(R)/VVI(R) mode,
for the dual and triple-chamber device in DDD(R) mode
•
Standard program with both high and low pacing energy
•
Data of the battery manufacturer (see the battery data)
Magnet
response
Cycles 1 to 10
After 10th cycle
Automatic
Asynchronous with 80 bpm
Synchronous with basic rate
reduced by 11%
Asynchronous
Asynchronous with 80 bpm
Asynchronous with 80 bpm
Synchronous
Synchronous with basic rate
reduced by 11%
Synchronous with basic rate
reduced by 11%
110 bpm
4.6 V
1.5 ms
500 Ω
30 bpm
0.2 V
0.1 ms
500 Ω
70 bpm
2.5 V
0.4 ms
500 Ω
70 bpm
5.0 V
0.4 ms
500 Ω
60 bpm
2.5 V
0.4 ms
600 Ω
60 bpm
5 V
0.4 ms
600 Ω
Mean value: 8 months
Minimum value: 6 months
—
Minimum value: 6 months
—
Minimum value: 6 months