30
Parameters
Pacing and Sensing
Pacing and Sensing
Pulse amplitude and
pulse width
Sensitivity
Atrial capture control
Ventricular capture control
Atrial overdrive pacing
Parameter
Range of values
Standard
SR
DR HF
Pulse
amplitude A/RV/LV
0.2 ... (0.2) ... 6.0 ... (0.5)
... 7.5 V
3.0 V
x
x
x
Pulse width A/RV/LV
0.1 ...(0.1) ... 0.5 ... (0.25)
... 1.5 ms
0.4 ms
x
x
x
Parameter
Range of values
Standard
SR
DR HF
Sensitivity
AUTO; 0.5 ... (0.5) ... 7.5 mV
AUTO
x
Sensitivity A
AUTO; 0.1 ... (0.1) ... 1.5 ...
(0.5) ... 7.5 mV
AUTO
x
x
RV sensitivity
AUTO; 0.5 ... (0.5) ... 7.5 mV
AUTO
x
x
x
LV sensitivity
OFF; AUTO; 0.5 ... (0.5) ...
7.5 mV
AUTO
x
Parameter
Range of values
Standard
SR
DR HF
Atrial capture control ATM (monitoring only); ON;
OFF
ON
x
x
Minimum amplitude
0.5 ... (0.1) ... 4.8 V
1.0 V
x
x
Threshold test start
2.4 ... (0.6) ... 4.8 V
3.0 V
x
x
Safety margin
0.5 ... (0.1) ... 1.2 V
1.0 V
x
x
Search type
Interval; time of day
Time of day
x
x
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
24 h
x
x
Time of day
00:00 ... (00:10) ...
23:50 hh:mm
00:30 hh:mm
x
x
Parameter
Range of values
Standard
SR
DR HF
Capture control RV
ATM (monitoring only); ON;
OFF
ON
x
x
x
Capture control LV
x
Minimum
amplitude RV
0.7 V
0.7 V
x
x
x
Minimum
amplitude LV
x
Threshold test start
2.4 ... (0.6) ... 4.8 V
3.0 V
x
x
x
RV safety margin
0.3 ... (0.1) ... 1.2 V
0.5 V
x
x
LV safety margin
1.0; 1.2 V
1.0 V
x
Search type
Interval; time of day
Time of day
x
x
x
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
24 h
x
x
x
Time of day
00:00 ... (00:10)
... 23:50 hh:mm
00:30 hh:mm x
x
x
Parameter
Range of values
Standard
SR
DR HF
Atrial overdrive
pacing
OFF; ON
With ON: maximum over-
pacing rate 120 bpm, mean
rate increase approximately
8 bpm, rate decrease after
20 cycles
OFF
x
x