PowerWave Operator’s Manual
3-11
•
Use SOFT KEY 1 to select the total reads option.
•
Use SOFT KEY 3 to select the duration option.
KINETIC INTERVAL ENTRY
Use this screen (Figure 3-12) to enter the interval of time time (in hours,
minutes and seconds) between each kinetic read.
K I N E T I C
I N T E R V A L :
0 1 : 2 3 : 5 6
Figure 3-12:
Kinetic Interval entry screen
•
Use the NUMERIC keys to enter the time duration. Valid ranges
are: 0-1 hours, 0-59 minutes and 0-59 seconds. The number of Reads
= Duration / Interval must be less than or equal to 40 and more
than or equal to 2.
•
Use the LEFT and RIGHT ARROW keys to move to the next or
previous numeric entry fields.