2-16
Installation
•
The fuse module has two fuse configurations:
The USA configuration has a fused Hot (PN 46023)
.75 amp slo blo).
The
European
configuration has both Hot and Neutral
fused (PN 46051) .315 amp slo blo T 5 X 20 mm.
•
To replace a bad fuse, pop out the old fuse and replace it with the
correct new one.
•
The configuration of the fuse module (U.S. or European) is
determined by which fusing network is facing the inside of the power
input module. To change the configuration, remove the Phillips head
screw that anchors the fuse holder. Remove the fuse holder and turn
it over. Replace the screw and reinstall the fuse module in the power
input module.