2-10
Installation
Changing the Baud Rate on the Reader
To change the baud rate from the default of 2400, to either 1200 or 9600:
•
Power up the instrument. At the Main Menu screen (Figure
2-8
), press the soft key beneath the menu option UTIL.
R E A D Y
9 : 4 5 P M
0 5 / 0 9 / 9 5
R E A D
D E F I N E R E P O R T U T I L
Figure 2-8
: The Main Menu screen
•
At the UTILITY OPTIONS screen, press the soft key beneath the
menu option SETUP (Figure 2-9).
S E L E C T U T I L I T Y O P T I O N ?
T E S T S
S E T U P
O U T P U T R E A D
Figure 2-9: Selecting Setup
•
At the EDIT SETUP INFORMATION screen, press the soft key
beneath the *MORE menu option to advance to the EDIT
SETUP
/ RS232 option screen (Figure 2-10).
E D I T
S E T U P I N F O R M A T I O N
D A T E
T I M E L A M B D A
* M O R E
E D I T
S E T U P I N F O R M A T I O N ?
R S 2 3 2
C A L P L A T E
* M O R E
Figure 2-10: Select *MORE to advance to the RS232 menu
Summary of Contents for PowerWave
Page 1: ...PowerWaveTM Microplate Scanning Spectrophotometer Operator s Manual ...
Page 2: ... ...
Page 4: ......
Page 6: ...ii ...
Page 18: ...xiv ...
Page 24: ...xx ...
Page 91: ...PowerWave Operator s Manual 3 39 Figure 3 49A Panel Report ...
Page 168: ...C 8 Error Codes ...
Page 170: ...D 2 Report Format Figure 1 Samples with Calls on Matrix Report ...
Page 171: ...Appendix D D 3 Figure 2 Curve Fit Report ...
Page 172: ...D 4 Report Format Figure 3 Samples with Calls on Column Report ...
Page 173: ...Appendix D D 5 Figure 4 Column Report without Samples ...
Page 174: ...D 6 Report Format Figure 5 Panel Report ...
Page 175: ...Appendix D D 7 Figure 6 Assay Detail Report ...