6

Warnings and Cautions

A.

Signal Words

The following signal words, as defined by ANSI Z535.4-1998, are used in the Toxi PlusOperator’s
Guide.

 indicates an imminently hazardous situation which, if not avoided, will result

in death or serious injury.

 indicates a potentially hazardous situation which, if not avoided, could result

in death or serious injury.

 indicates a potentially hazardous situation, which if not avoided, may result

in moderate or minor injury.

CAUTION

 used without the safety alert symbol indicates a potentially hazardous situation

which, if not avoided, may result in property damage.

B.

Warnings

1. 

 The Toxi Plus personal, portable gas detector has been designed for the

detection of specific toxic gas accumulations.  An alarm condition indicates the presence of a
potentially life-threatening hazard and should be taken very seriously.

2.

  In the event of an alarm condition it is important to follow established

procedures.  The safest course of action is to immediately leave the affected area, and to return only
after further testing determines that the area is once again safe for entry.  Failure to immediately
leave the area may result in serious injury or death.

3.

 Hand aspirated remote sampling only provides continuous gas readings as

long as the bulb is being squeezed.

4.

 The accuracy of Toxi Plus instruments should be checked periodically with

known concentration calibration gas.  Failure to check accuracy can lead to inaccurate and
potentially dangerous readings.

5.

 A sensor that cannot be calibrated or is found to be out of tolerance must be

replaced immediately.  An instrument equipped with a sensor that fails calibration may not be used
until testing with known concentration test gas determines that accuracy has been restored, and the
instrument is once again fit for use.

6.

  Do not reset the calibration gas concentration setpoints in the Toxi Plus

unless the concentration of your calibration gas differs from the concentration of the calibration gas
that is normally supplied by Biosystems for use in calibrating the Toxi Plus.

7.

  Use of non-standard calibration gas and/or calibration kit components when

calibrating the Toxi Plus can lead to dangerously inaccurate readings and may void the standard
Biosystems warranty.
Biosystems offers calibration kits and long-lasting cylinders of test gas specifically developed for
easy calibration.  Customers are strongly urged to use only Biosystems calibration materials when
calibrating the Toxi Plus

8.

  Substitution of components may impair intrinsic safety.

9.

  For safety reasons the Toxi Plus must be operated by qualified personnel

only.  Read, understand and follow the directions set forth in this reference manual before operating
the Toxi Plus.

10.

  The Toxi Plus has been tested for intrinsic safety in Explosive Gas/AIR(21.0%

O2) Only.

Summary of Contents for Toxi Plus

Page 1: ...0 Reference Manual Toxi Plus Single Sensor Gas Detector 651 South Main St Middletown CT 06457 860 344 1079 FAX 860 344 1068 04AUG2004 P N 13 044 Version 3 3...

Page 2: ...1...

Page 3: ...ONS IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED AVERTISSEMENT LIRE ATTENTIVEMENT LES INSTRUCTIONS AVANT DE METTRE EN MARCHE Toxi Plus Operation Manual Co...

Page 4: ...mote sample draw assembly 10 1 9 4 Weather cover 10 1 9 5 Training video 10 1 10 Toxi Plus design components 10 1 11 Toxi Plus accessories 11 Chapter 2 Basic operation 12 2 1 Overview of Toxi Plus ope...

Page 5: ...Re setting the calibration gas concentration 25 4 3 Setting the security beep interval 26 Chapter 5 Trouble Shooting and Repair 27 5 1 Replacing Toxi Plus sensor 27 5 2 Troubleshooting 27 5 2 1 Unit w...

Page 6: ...to day operation including fully automatic calibration adjustment through use of the single on off Mode button The Toxi Plus is Classified by Underwriters Laboratories Inc and the Canadian Standards...

Page 7: ...oxi Plus instruments should be checked periodically with known concentration calibration gas Failure to check accuracy can lead to inaccurate and potentially dangerous readings 5 A sensor that cannot...

Page 8: ...7...

Page 9: ...us is designed to detect is determined by the sensor The type of sensor must be specified at the time the instrument is purchased The types of sensors currently available include substance specific el...

Page 10: ...With the alarms latched once an alarm occurs both visual and audible alarms continue to sound even after the atmospheric hazard has cleared The instrument must be manually reset to silence the alarms...

Page 11: ...nd H2S The sensor selected must be specified at the time of purchase 1 9 2 Vibrator alarm An internally mounted vibrating alarm is optionally available for use in high noise or other environments requ...

Page 12: ...trols all normal day to day operations including Auto Calibration adjustment Accurate electrochemical sensor Easy to read LCD display with backlight Loud two tone audible alarm Optional built in vibra...

Page 13: ...ator alarm will briefly activate During start up the LCD display will also show several messages or screens in sequence The first screen displays the software version number Note The software version...

Page 14: ...ormal operation of the instrument resumes 2 2 3 Technician mode Technician Mode provides access to all advanced functions and displays When the instrument is turned on in the Technician Mode following...

Page 15: ...nging operating modes It s possible to change operating modes while the instrument is in use without affecting the on going calculations used to determine Peak STEL and TWA alarms Simply slide the bel...

Page 16: ...moisture barrier filter Normal air movements are enough to carry the sample to the sensor The sensor reacts to changes in the concentration of the gas being measured Values are quickly updated and dis...

Page 17: ...kits and long lasting cylinders of test gas specifically developed for easy Toxi Plus calibration Customers are urged to use Biosystems calibration materials when calibrating their Toxi Plus detector...

Page 18: ...ach is usually to use a direct reading substance specific sensor If hydrogen sulfide is known to be potentially present the sensor selected should be specifically for the detection of H2S and calibrat...

Page 19: ...esh air If the Toxi Plus is operated in either the Basic or Technician mode readings should equal 0 PPM If the instrument is operated in the Text Only mode all readings should indicate conditions are...

Page 20: ...initiating auto calibration procedures 3 Make sure the instrument is located in an area where the air is known to be fresh 4 Press the MODE button 3 times within two seconds This will wake up the inst...

Page 21: ...as it is screwed into the cylinder of gas 3 The instrument will display the sensor readings as they rise When they have stabilized at their highest value the instrument will note the response and if...

Page 22: ...ly which verifies that the zero values have not been changed The instrument will then return to normal operation 3 5 2 Span calibration using keypad buttons Span calibration procedures using buttons o...

Page 23: ...it may indicate that the sensor is unresponsive or dead and requires replacement 11 When span calibration has been completed press and hold the Cal button down until the LCD screen indicates that cal...

Page 24: ...o sound even after the atmospheric hazard has cleared Press the MODE button to reset the instrument silence the alarms and restore normal operation Procedures for latching Toxi Plus alarms are given i...

Page 25: ...ode as discussed in Section 2 2 4 then turn the instrument off 2 Turn the instrument on while pressing and holding down the Alarm button This will place the instrument in the Alarm Adjust Mode 3 Adjus...

Page 26: ...ry When gas is applied to the sensor during auto calibration the instrument automatically adjusts the sensitivity of the carbon monoxide to produce readings that match this 50 PPM concentration Sensor...

Page 27: ...ocedure to turn the security beep on 1 Make sure the Toxi Plus detector is in the Technician mode as discussed in Section 2 2 4 then turn the instrument off 2 Turn the instrument on while pressing and...

Page 28: ...g and repair procedures that can be done in the field Caution The following repair procedures should only be performed by authorized personnel 5 2 1 Unit will not turn on Possible causes Battery disch...

Page 29: ...ny calibration gas fittings take the instrument to fresh air and allow readings to stabilize Do a manual fresh air zero adjustment using buttons on the instrument keypad as discussed in Section 3 5 1...

Page 30: ...ood example of an acutely toxic substance which is immediately lethal at relatively low concentrations Exposure to a 1 000 PPM parts per million concentration of H2S in air produces rapid paralysis of...

Page 31: ...osure limits which are higher than the eight hour TWA The STEL is the maximum average concentration to which an unprotected worker may be exposed in any fifteen minute interval during the day During t...

Page 32: ...rms should be set If OSHA has not determined a ceiling value for greatest safety the Toxi Plus ceiling alarm should be set at the same value as the STEL alarm If OSHA has not determined a STEL value t...

Page 33: ...alibration easy in this case it is necessary to use special calibration Zero Contaminant gas This gas cylinder Biosystems part number 9039 is used in conjunction with the sample draw calibration adapt...

Page 34: ...H2S Sulfur dioxide SO2 CO 100 10 100 25 0 H2S 10 100 350 100 0 SO2 10 20 65 15 100 NO 30 0 25 6 0 NO2 15 20 60 15 120 Cl2 10 20 40 10 5 H2 60 5 60 15 0 HCN 15 0 40 10 50 HCl 3 0 5 1 0 NH3 0 0 0 Ethy...

Page 35: ...e instrument is in service Toxi Oxy Pro 2 years from date of purchase Toxi Oxy Limited 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Mighty Tox 90 days...

Reviews: