background image

44

If a location has not been assigned, the screen will
be blank other than showing “ID=” 

 or 

Follow the directions above for the user name and 
enter the location name.  Once the location has 
been entered, press the mode button to save the 
monitoring location and return to normal operation. 

Note:  Only the current user name and 
monitoring location can be modified through the 
instrument keypad.  Modification of the user 
name and location lists in instrument memory 
requires use of Biosystems’ Datalink Software 
and connection with a personal computer as 
discussed in the PhD Ultra Datalink Manual. 

4.5.6  Downloading recorded data to a 

computer 

PhD Ultra record keeping capabilities are most 
useful when used together with Biosystems Datalink 
Gas Detector Database Software for use with IBM 
compatible Personal Computers.  To download data 
to a personal computer: 

(1)  Make sure the PhD Ultra is turned off.   

(2)  Connect the PhD Ultra Datadock RS-232 cable 

to an available COM port on your computer. 

(3)  Slip the PhD Ultra into the Datadock. 

(4)  Load Biosystems PhD Ultra Datalink software 

on your personal computer. 

(5)  Choose “Retrieve Data from Instrument” from 

the “Datalogger” menu. 

(6)  The Datalink software will automatically “wake 

up” the PhD Ultra, and initiate data transfer. 

(7)  When downloading is complete the PhD Ultra 

will shut itself back off.

Consult the Datalink Owner’s Manual for instructions 
on how to review, store and export data, generate 
reports, or make use of other advanced features for 
the examination and storage of data. 

4.5.7  Display “Service Due” dates 

PhD Ultra Datalink software may be used to assign 
an optional “Service Due” date for each sensor.  Any 
time after this date is passed a “Service Due” 
message will be displayed whenever the instrument 
is first turned on.  Pressing the mode button 
acknowledges the message and allows normal 
operation.  The message will continue to be 
displayed each time the instrument is turned on until 
a new service due date is assigned.  Since this 
information is stored directly with the sensor 
EEPROM, a service due alarm will be activated 
even if the sensor is removed from one instrument 
and installed in another.   

Service due dates may be reviewed either via 
Datalink software or by using buttons on the 
instrument keypad to display the dates directly on 
the instrument LCD. 

Note:  It is not possible to assign a new service 
due date through the instrument keypad.  
Changing or assigning a new service due date 
may only by done via personal computer and 
PhD Ultra Datalink software.   

Service due dates may be reviewed by pressing the 

-

” button at any time while the instrument is being 

used in a normal operating mode.  A screen will
indicate as the instrument begins to display the 
assigned dates.   

This screen will be followed others showing the 
assigned service due date for each sensor currently 
installed.  These screens will continue to be shown 
in rotation. 

p

p

etc. 

If no service due date has been assigned, the 
screen will indicate that the warning has been
disabled.

 

Pressing the mode button at any time returns the 
instrument to normal operation. 

4.6 Passcode 

Overview 

The PhD Ultra has an option that when enabled, 
requires that a passcode be entered before any of 
the advanced features or the automatic span 
calibration can be activated. This passcode is a 4 
digit number which can be set to any number from 
“0 0 0 1” to “9 9 9 9”.  

To enter passcode options, turn the PhD Ultra on 
while holding down both the “+” and “ALM” buttons. 
When you hear an audible beep from the alarm let 
go of the “+” and “ALM” buttons. The Setup 
Passcode screen will be shown briefly followed by 
the Enter Passcode screen. 

Summary of Contents for PhD Ultra

Page 1: ...Reference Manual PhD Ultra Multi Gas Detector 651 South Main Street Middletown CT 06457 USA 860 344 1079 800 711 6776 FAX 860 344 1068 http www biosystems com Version 2 60 28JAN2008 Part Number 13 03...

Page 2: ...NS IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED AVERTISSEMENT LIRE ATTENTIVEMENT LES INSTRUCTIONS AVANT DE METTRE EN MARCHE PhD Ultra Operation Manual Ver...

Page 3: ...6 TEMPERATURE OUT OF RANGE 12 1 6 4 7 OTHER ELECTRONIC SAFEGUARDS 12 1 6 4 8 SECURITY BEEP 12 1 7 INSTRUMENT FIRMWARE REQUIREMENT FOR COMPATIBILITY WITH THE IQ SYSTEM 12 1 8 CLASSIFICATION FOR INTRIN...

Page 4: ...2 4 3 2 CHANGING SAMPLE PROBE TUBES 23 2 5 BIOSYSTEMS EEPROM EQUIPPED SMART SENSORS 24 2 5 1 Identification of type of sensor by instrument 24 2 5 2 Other information stored with the sensor EEPROM 24...

Page 5: ...38 4 4 RE INITIALIZING THE PHD ULTRA 38 4 5 RECORD KEEPING 39 4 5 1 PhD Ultra datalogging overview 39 4 5 2 Optional Datalink and Gas Detection Database Software Kit 39 4 5 3 Adjusting record keeping...

Page 6: ...IZED PUMP MAINTENANCE 49 5 4 1 Internal pump filter replacement 50 5 4 2 Specific problems with motorized pump 51 5 4 2 1 PUMP WILL NOT TURN ON 51 5 4 2 2 CAN T RESUME NORMAL OPERATION AFTER A LOW FLO...

Page 7: ...for combustible gas monitoring in atmospheres where oxygen concentrations exceed 21 0 oxygen ONLY THE COMBUSTIBLE GAS DETECTION PORTION OF THIS INSTRUMENT HAS BEEN ASSESSED FOR PERFORMANCE BY CSA UNIQ...

Page 8: ...An instrument that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrument is once again fit for use 7 Do not r...

Page 9: ...ectrochemical toxic sensors that have been designed to minimize the effects of common interfering gases These sensors provide accurate dependable readings for toxic gases commonly encountered during c...

Page 10: ...c reading may indicate a hazardous combustible gas concentration that exceeds the PhD Ultra s zero to 100 percent LEL detection range Failure to immediately leave the area may result in serious injury...

Page 11: ...potentially life threatening atmospheric conditions Any alarm condition should be taken seriously The safest course of action is to immediately leave the affected area and return only after further te...

Page 12: ...nt firmware version is given immediately after the PhD Ultra is turned on 1 8 Classification for intrinsic safety The PhD Ultra is Classified by Underwriters Laboratories Inc and the Canadian Standard...

Page 13: ...nt is being used in a hazardous location Alkaline battery packs may not be opened and alkaline batteries may not be replaced while the battery pack is located in a hazardous area Battery replacement a...

Page 14: ...he padded leather weather cover 11 Key pad The key pad is located on the back of the PhD Ultra under the belt clip Slide the belt clip towards the rear of the instrument to access the four small push...

Page 15: ...ill then be shown Note If the date and time are incorrect see section 4 5 3 3 for instructions The PhD Ultra automatically evaluates itself to determine its electronic fitness for use by performing an...

Page 16: ...be put back into service or used until the accuracy of any affected sensor has been verified by exposure to the appropriate known concentration test gas The Needs Cal warning message may be acknowled...

Page 17: ...cal reading changes to reflect the new value the LED alarm light flashes and the audible alarm sounds PhD Ultra alarms are normally self resetting When readings drop below the pre set alarm levels vis...

Page 18: ...instrument keypad Push the and buttons at the same time to change operating modes Each time the operating mode is changed the display screen will briefly indicate the operating mode that has been sel...

Page 19: ...rinsic safety 2 3 3 Low battery alarms The PhD Ultra is designed to automatically determine which type of battery pack has been installed and use the appropriate low battery alarms Alarms will be acti...

Page 20: ...in and check to see that the PWR power on indicator LED on the charger cradle is lit 4 Slip the PhD Ultra into the charger cradle and check to see that the FAST fast charger indicator LED on the charg...

Page 21: ...ample draw adapter 3 Connect the other end of the hose to the sample probe as shown in Figure 2 4 1 1 4 Attach the sample draw adapter to the PhD Ultra as shown in Figure 2 4 1 2 5 Cover the end of th...

Page 22: ...4 2 1 Using the continuous sample draw pump 1 Turn the PhD Ultra on Figure 2 4 2 1 1 Sample draw pump and probe assembly 2 Connect the slip on battery operated pump with the hose and probe assembly as...

Page 23: ...taminants are removed by means of a cellulose filter similar to those used in filter cigarettes The hydrophobic filter includes a 0 1 Pm Teflon barrier which blocks the flow of moisture as well as any...

Page 24: ...henever a change is made during operation and whenever the instrument is turned off If a sensor is changed or replaced the PhD Ultra notes that a change has occurred displays a Needs Cal message the n...

Page 25: ...to inaccurate and potentially dangerous readings and may void the standard Biosystems warranty Customers are strongly urged to use only Biosystems calibration materials when calibrating the PhD Ultra...

Page 26: ...in serious injury or death Avertissement Toute lecture rapide et positive suivie d une baisse subite au erratique de la valeur peut indiquer une concentration de gaz hors gamme de d tection qui peut...

Page 27: ...instrument will display H2S and hydrogen sulfide must be used to verify accuracy Use of the incorrect calibration gas may lead to inaccurate and potentially dangerous readings If the sensor is calibr...

Page 28: ...d value during a bump test fails to fall between 100 and 120 of the expected value for the gas Always check the expiration date on calibration gas cylinder s prior to use Expired calibration gas can l...

Page 29: ...Calibration Adjustment screen This screen also includes a timer that counts down the number of seconds left to initiate the adjustment Note If the MODE button is pressed at any time prior to completi...

Page 30: ...en the instrument has detected that the proper gas has been applied the sensor will be adjusted 6 When all sensors currently installed have been successfully span adjusted the display will announce Au...

Page 31: ...ns Span calibration procedures using buttons on the instrument keypad are most useful when calibration of only a single sensor is desired Span calibration using the keypad buttons may only be done whi...

Page 32: ...gas is applied to the sensor being adjusted a screen will be displayed indicating that the span gas concentration is too low Verify that the flow rate of the regulator is 1 0 liters per minute Replac...

Page 33: ...tings can be restored at any time during normal operation by using the procedures discussed in Section 4 2 2 In many cases it is possible to comply with OSHA guidelines while using higher alarm points...

Page 34: ...reens would be shown Oxygen and combustible gas LEL The ceiling alarm set points for the toxic sensors currently installed The STEL and TWA alarm set points for the toxic sensor currently installed in...

Page 35: ...L button to advance the display to the next available setup option or press Mode to exit from the instrument setup screens 4 3 2 Changing the precision of the toxic sensor read out The toxic decimal p...

Page 36: ...age showing the minutes and seconds between beeps Use the and keys to set the interval Setting the interval to 0 will turn the security beep off 4 3 7 Low temperature alarms To access the low temperat...

Page 37: ...on Gas screen If an LEL combustible gas sensor has been installed the LEL sensor s calibration gas concentrations will be the first shown This screen indicates the concentration and type of test gas t...

Page 38: ...3 seconds A screen will briefly announce that changes have been made and the instrument setup has been updated The PhD Ultra will then turn itself off Pressing mode button to cancel the changes and s...

Page 39: ...ing data may be stored Datalogging is a transparent function that is continually in operation As long as the datalogger has not been disabled the instrument will begin recording data as soon as the in...

Page 40: ...o not effect the way in which TWA STEL Ceiling and Peak exposure values are calculated To modify the datalogger interval first enter the datalogging adjust mode as described in section 4 5 3 1 Once th...

Page 41: ...he software is set to the other proper communication will not be possible The software communication rate setting can be modified through the PC Setup screen as discussed in the PhD Ultra Datalink man...

Page 42: ...sion Press to view the oldest monitoring session in the datalogger memory After a few seconds the PhD Ultra will automatically show the most recent monitoring session The numbers in this screen signif...

Page 43: ...user ID is set press the mode button to display the list of locations currently in the instrument memory The first screen will show the current monitoring location on the top row with the name of the...

Page 44: ...be displayed whenever the instrument is first turned on Pressing the mode button acknowledges the message and allows normal operation The message will continue to be displayed each time the instrumen...

Page 45: ...e current status of the passcode option To disable the passcode option press the button To enable the passcode option press the button Press the CAL button to save the new setting Press the mode butto...

Page 46: ...lunt tool B4 Remove sensor blank from the sensor compartment Then proceed to step C5 or D5 depending on the sensor type For Sensors O2 LEL CO CO Plus H2S NO perform step C5 C5 Place the new filter sna...

Page 47: ...nt is connected to a 110 VAC power source through the battery charger The PhD Ultra must be located in a non hazardous location during the charging cycle Charging the PhD Ultra in a hazardous location...

Page 48: ...Technician operating mode Take the instrument to a non hazardous location If equipped with an alkaline battery pack replace the batteries and attempt to turn the instrument back on If equipped with a...

Page 49: ...en supplied with your replacement PROM Do not use small screwdrivers or other non recommended devices to remove the old PROM Use of non recommended devices can easily cause damage to the PROM socket C...

Page 50: ...sections of the pump case as shown in Figure 5 4 3 2 Remove the two pieces of tubing from the filter cap by gently twisting and pulling until the ends are free 3 Remove the filter cap and replace the...

Page 51: ...5 5 Returning your PhD Ultra to Biosystems for service or repair Please contact the Biosystems Service Department at 860 344 1079 to obtain a Return Authorization number prior to shipment A Biosystems...

Page 52: ...rt oxygen Although very high concentrations of carbon monoxide may be acutely toxic and lead to immediate respiratory arrest or death it is the long term physiological effects due to chronic exposure...

Page 53: ...to be present in LEL concentrations in order for this reaction to occur Even trace amounts of combustible gas present in the air surrounding the sensor will be catalytically burned on the surface of t...

Page 54: ...1910 146 OSHA has determined that a combustible hazard exists whenever the concentration of combustible gas or vapor exceeds 10 percent LEL Per this standard confined spaces with concentrations which...

Page 55: ...inst this potentially dangerous sort of calibration error Biosystems Propane Equivalent calibration mixtures are based on methane but in concentrations which are designed to produce a level of sensiti...

Page 56: ...ppm 1 0 ppm HCN 4 7 ppm 4 7 ppm 4 7 ppm NH3 25 ppm 25 ppm 35 ppm NO 25 ppm 25 ppm 25 ppm NO2 1 0 ppm 1 0 ppm 1 0 ppm Oxygen O2 Low alarm 19 5 High alarm 22 0 Combustible gas 10 LEL Biosystems PhD Ult...

Page 57: ...EZ Cal Value Pack multi component calibration gas CO 50 ppm propane equivalent 50 LEL in air Only for use with 54 12 022 Value Pack style fixed flow rate regulator 54 9041 EZ Cal multi component gas...

Page 58: ...s has with the sensor the effect can either decrease the signal negative cross sensitivity or increase the signal positive cross sensitivity Each figure represents the reaction of the sensor to 100 pp...

Page 59: ...closely follow these guidelines If your operating procedures do not permit daily checking of the sensors Biosystems recommends the following procedure to establish a safe and prudent accuracy check s...

Page 60: ...ears from date of purchase Toxi Limited 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Toxi3Ltd 3 years after activation or 3 years after the Must Be Act...

Reviews: