Biosan UVR-M User Instructions Download Page 8

 

Figure 6. Lamp runtime 

5.2.7 

End of lamp lifetime and lamp failure

. At the end of lamp lifetime (9000 h) or in 

case of UV lamp failure, unit terminates operation and does not resume until lamps 
are replaced. Display shows “END OF UV LAMP LIFETIME” or “ERROR UV LAMP 
FAILURE” messages, respectively. Additionally, check the indicator (fig. 5/8) to see 
if the lamps are working. To replace the lamps, see 

8.3

5.2.8  After finishing the operation, switch off the power switch and unplug the power cable 

from the mains. 

6. 

Specifications 

The unit is designed for operation in cold rooms and closed laboratory rooms at am-

bient temperature from +4°C to +40°C in a non-condensing atmosphere and maximum rel-
ative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity 
at 40°C. 

Biosan is committed to a continuous programme of improvement and reserves the 

right to alter design and specifications of the equipment without additional notice. 

Model 

UVR-M 

UVR-Mi 

UV radiation source 

Low ozone, bactericidal, TUV 25W G13 UV-C 

1 lamp 

2 lamps 

Wavelength 

253.7 nm 

Airflow speed 

21 m

3

/h 

29 m

3

/h 

Protection from direct UV light 

Full 

Lamp service life  

9000 h 

Lamp operation indicator 

Yes 

Display 

No 

LCD, 2x16 characters 

Lamp service life counter 

No 

Yes 

Lamp fault detection 

No 

Yes 

Timer 

No 

1 min - 24 h / non-stop 

Clock and on/off scheduler 

No 

Yes 

Dimensions  

130x110x660 mm 

Weight, a

ccurate within ±10% 

2.74 kg 

2.38 kg 

Nominal operating voltage, frequency  220 V~, 50/60 Hz 

/ 120V~, 60 Hz

 

220 V~, 50/60 Hz 

Power consumption 

125 W (0.54 A) 

/ 160 W (1.3 A)

 

110 W (0.5 A) 

 

 

 

Summary of Contents for UVR-M

Page 1: ......

Page 2: ...services we would like to hear from you Please send all feedback to Manufacturer SIA Biosan Ratsupites iela 7 k 2 Riga LV 1067 Latvia Phone 371 674 261 37 Fax 371 674 281 01 https biosan lv Marketing...

Page 3: ...special attention to sections marked by this symbol Caution Do not switch on the unit with the cover removed or without filters UV lamp must be covered at all times during operation Otherwise UV lamp...

Page 4: ...specific at mospheres Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms 2 4 Biological safety The user is responsible to carry out appropri...

Page 5: ...t damage Warranty covers only the units transported in the original package 4 2 Complete set Package contents 4 2 1 Standard set UVR M UVR Mi UV Cleaner Recirculator 1 pce Spare dust filters 2 pcs Spa...

Page 6: ...ta given by the manufacturer so lamp replacement or air recircu lation time extension by 12 is recommended in order to achieve the de sired air cleaning result Figure 4 UVR M overview 5 1 Working with...

Page 7: ...the next parameter 5 2 3 3If no key is pressed for more than 10 s the changes will be saved 5 2 4 The unit can operate in automatic or manual modes 5 2 4 1Manual mode Press the Run Stop key fig 5 7 to...

Page 8: ...nd maximum rel ative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C Biosan is committed to a continuous programme of improvement and reserves the right to...

Page 9: ...rating integrity check If the unit follow the procedure described in section 5 Op eration then no additional checks are required 8 2 Cleaning and disinfection 8 2 1 Use mild soap and water with a soft...

Page 10: ...y Medium 8 5 Decommissioning of the unit The user is responsible for the decontamination of the unit before decommissioning Dispose of the appliance as electronic equipment in accordance with the rele...

Page 11: ...and retain for your records Model Serial number Date of sale UVR M UVR Mi UV Cleaner Recirculator 10 8 Production date Production date is placed in the serial number on the label of the unit Serial n...

Page 12: ...SIA Biosan R tsup tes iela 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 https biosan lv Edition 3 4 01 October of 2021...

Reviews: