Biosan UVR-M User Instructions Download Page 6

4.4 

Effective air recirculation

. Table below lists room area and volume where air effec-

tively recirculated by one unit. 

Model 

UVR-M 

UVR-Mi 

Room with active ventilation  

(working air conditioner or air hood) 

10 square metres /  

30 cubic metres 

15 square metres /  

45 cubic metres 

Ordinary ventilated room 

20 square metres /  

60 cubic metres  

30 square metres /  

90 cubic metres 

 

Note. 

For larger rooms, increase unit count or type accordingly. 

5. 

Operation 

 

Caution! 

Bactericidal lamp properties reduce by ~12% after 5000 hours of opera-
tion (data given by the manufacturer), so lamp replacement or air recircu-
lation time extension by 12% is recommended in order to achieve the de-
sired air cleaning result. 

 

Figure 4. UVR-M overview 

5.1 

Working with model UVR-M

5.1.1  Plug  the  power  cord  into  a  grounded  mains  socket.  Switch  on  the  power  switch 

(fig. 4/1). 

5.1.2  Unit starts the recirculator automatically. Lit indicator (fig. 4/2) shows that the UV lamp 

is working. 

5.1.3  After finishing the operation, switch off the power switch and unplug the power cable 

from the mains. 

5.1.4 

Lamp  failure

.  In  case  of  UV  lamp  failure,  unit  terminates  operation  and  does  not 

resume until the lamp is replaced. Check the indicator (fig. 4/2) to see if the lamp is 
working. To replace the lamp, see 

8.3

Summary of Contents for UVR-M

Page 1: ......

Page 2: ...services we would like to hear from you Please send all feedback to Manufacturer SIA Biosan Ratsupites iela 7 k 2 Riga LV 1067 Latvia Phone 371 674 261 37 Fax 371 674 281 01 https biosan lv Marketing...

Page 3: ...special attention to sections marked by this symbol Caution Do not switch on the unit with the cover removed or without filters UV lamp must be covered at all times during operation Otherwise UV lamp...

Page 4: ...specific at mospheres Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms 2 4 Biological safety The user is responsible to carry out appropri...

Page 5: ...t damage Warranty covers only the units transported in the original package 4 2 Complete set Package contents 4 2 1 Standard set UVR M UVR Mi UV Cleaner Recirculator 1 pce Spare dust filters 2 pcs Spa...

Page 6: ...ta given by the manufacturer so lamp replacement or air recircu lation time extension by 12 is recommended in order to achieve the de sired air cleaning result Figure 4 UVR M overview 5 1 Working with...

Page 7: ...the next parameter 5 2 3 3If no key is pressed for more than 10 s the changes will be saved 5 2 4 The unit can operate in automatic or manual modes 5 2 4 1Manual mode Press the Run Stop key fig 5 7 to...

Page 8: ...nd maximum rel ative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C Biosan is committed to a continuous programme of improvement and reserves the right to...

Page 9: ...rating integrity check If the unit follow the procedure described in section 5 Op eration then no additional checks are required 8 2 Cleaning and disinfection 8 2 1 Use mild soap and water with a soft...

Page 10: ...y Medium 8 5 Decommissioning of the unit The user is responsible for the decontamination of the unit before decommissioning Dispose of the appliance as electronic equipment in accordance with the rele...

Page 11: ...and retain for your records Model Serial number Date of sale UVR M UVR Mi UV Cleaner Recirculator 10 8 Production date Production date is placed in the serial number on the label of the unit Serial n...

Page 12: ...SIA Biosan R tsup tes iela 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 https biosan lv Edition 3 4 01 October of 2021...

Reviews: