Biosan UVR-M User Instructions Download Page 5

UV radiation affects viability of microorganisms by causing photochemical reactions 

in the structure of DNA and RNA. Adjacent pyrimidine molecules form dimers and block the 
reproduction of microorganisms, as a result, causing their death. The diagram on figure 2 
shows the process of formation of pyrimidine dimers, using thymine as an example. 

 

Figure 2. Photochemical reaction, the formation of pyrimidine dimers, 

thymine taken as an example  

(source http://www.photobiology.info) 

4. 

Getting started 

4.1 

Unpacking.

 Remove packing materials carefully and retain them for future shipment 

or  storage  of  the  unit.  Examine  the  unit  carefully  for  any  damage  incurred  during 
transit. The warranty does not cover in-transit damage. Warranty covers only the units 
transported in the original package. 

4.2 

Complete set.

 Package contents: 

4.2.1  Standard set: 

UVR-M

 / 

UVR-Mi

, UV Cleaner-Recirculator ..................................................... 1 pce. 

Spare dust filters .............................................................................................. 2 pcs. 

Spare fuse  ....................................................................................................... 1 pce. 

Screw with dowel ............................................................................................. 1 pce. 

User instructions, declaration of conformity  .................................................... 1 copy 

4.2.2  Optional accessories, on request: 

UVR-S movable support stand ........................................................................... 1 set 

4.3 

Setup

Choose recirculator location so that the air intake and discharge operate unim-
peded, providing the best air circulation in the room. 

 

Caution! 

UV recirculator must be used only in clean rooms! Dust accumulation on 
the electrical parts of the device can cause a short circuit. 

Fix the unit on the wall with the screw and dowel, or on a UVR-S movable support 
stand (figure 3). 

Position the unit so that the access to the mains plug is unobstructed. 

 

 

 

 

 

Attachment point 

Screw and dowel 

Attachment on UVR-S stand 

Figure 3. Unit attachment methods 

Summary of Contents for UVR-M

Page 1: ......

Page 2: ...services we would like to hear from you Please send all feedback to Manufacturer SIA Biosan Ratsupites iela 7 k 2 Riga LV 1067 Latvia Phone 371 674 261 37 Fax 371 674 281 01 https biosan lv Marketing...

Page 3: ...special attention to sections marked by this symbol Caution Do not switch on the unit with the cover removed or without filters UV lamp must be covered at all times during operation Otherwise UV lamp...

Page 4: ...specific at mospheres Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms 2 4 Biological safety The user is responsible to carry out appropri...

Page 5: ...t damage Warranty covers only the units transported in the original package 4 2 Complete set Package contents 4 2 1 Standard set UVR M UVR Mi UV Cleaner Recirculator 1 pce Spare dust filters 2 pcs Spa...

Page 6: ...ta given by the manufacturer so lamp replacement or air recircu lation time extension by 12 is recommended in order to achieve the de sired air cleaning result Figure 4 UVR M overview 5 1 Working with...

Page 7: ...the next parameter 5 2 3 3If no key is pressed for more than 10 s the changes will be saved 5 2 4 The unit can operate in automatic or manual modes 5 2 4 1Manual mode Press the Run Stop key fig 5 7 to...

Page 8: ...nd maximum rel ative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C Biosan is committed to a continuous programme of improvement and reserves the right to...

Page 9: ...rating integrity check If the unit follow the procedure described in section 5 Op eration then no additional checks are required 8 2 Cleaning and disinfection 8 2 1 Use mild soap and water with a soft...

Page 10: ...y Medium 8 5 Decommissioning of the unit The user is responsible for the decontamination of the unit before decommissioning Dispose of the appliance as electronic equipment in accordance with the rele...

Page 11: ...and retain for your records Model Serial number Date of sale UVR M UVR Mi UV Cleaner Recirculator 10 8 Production date Production date is placed in the serial number on the label of the unit Serial n...

Page 12: ...SIA Biosan R tsup tes iela 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 https biosan lv Edition 3 4 01 October of 2021...

Reviews: