2. General Information
Microplate Photometer
HiPo MPP-96
is a compact tabletop device for analysing the
results of ELISA and microbiological studies by measuring optical density in 96-well micro-
plates. Instrument is controlled by and outputs data to computer.
Scheme 1. Optical scheme of the unit
HiPo MPP-96
is supplied with analysis software QuantAssay. Software performs a
single scanning with option of a second scanning on reference channel available. When
using QuantAssay, it is possible to program the analysis of the following assays:
Quantitative assays: the ability to install up 20 standards and choose fit model from
5/4 parameter logistic, linear and piecewise linear models.
BestFit function for the selection of the best calibration curve.
Multiplex analysis - up to 7 different tests on the same plate.
Qualitative assays: the ability to install up to 8 types of controls (weak positive, strong
positive, negative, etc.).
Avidity / affinity assay.
Save, load and export results.
Create visual reports.
Together with standard light filters (with wavelengths of 405, 450, 492, 620 nm), it is
possible to order light filters in the range from 400 to 700 nm, installed by manufacturer. See
the list on page 12 in the
Specifications
section.
HiPo MPP-96
is factory calibrated using Thermo Scientific Multiskan Verification
Plate with 8 wavelengths. Certificate of verification is included with the unit.
8 Si photodiodes
8 apertures
8-well row #0,
for self-calibration of OD
8-channel optical fiber
White LED (430-700 nm)
Purple LED (400-430 nm)
Mechanical Mirror
Filter wheel with 8 places
Lens