1.  Safety Precautions 

The following symbol means: 

 

Caution:

 

Make sure you have fully read and understood the present  Manual 
before using the equipment. Please pay special attention to sections 
marked by this symbol. 

GENERAL SAFETY 

 

Use only as specified in the operating manual provided. 

 

Save the unit from shocks and falling.  

 

Store and transport the unit at ambient temperatures between -20

°C and +60°C and 

maximum relative humidity of 80%. 

 

After  transportation  or  storage,  keep  the  unit  under  room  temperature  for  2  to  3 h 
before connecting it to the mains.  

 

Before using any cleaning or decontamination methods except those recommended 
by the manufacturer, check with the manufacturer that the proposed method will not 
damage the equipment. 

 

Do not make modifications in design of the unit.  

 

Do not place the optic parts under direct light. 

 
ELECTRICAL SAFETY 

 

Connect only to an external power supply with voltage corresponding to that on the 
serial number label. 

 

Use only the external power supply provided with this product. 

 

Ensure that the power switch and external power supply are easily accessible during 
use. 

 

Disconnect the unit from the external power supply before moving. 

 

This equipment is connected and controlled by PC. Please ensure that the attached 
PC itself conforms to safety and EMC standards. 

 

If  liquid  penetrates  into  the  unit,  disconnect  it  from  the  external  power  supply  and 
have it checked by a repair and maintenance technician. 

 

Do not operate the unit in premises where condensation can form. Operating condi-
tions of the unit are defined in the Specifications section. 

 
DURING OPERATION  

 

Do not operate the unit in environments with aggressive or explosive chemical mix-
tures. Please contact manufacturer for possible operation of the unit in specific at-
mospheres. 

 

Do not operate the unit if it is faulty or was installed incorrectly. 

 

Do not use outside laboratory rooms.  

 

Set up the PC display with the correct viewing position, free from glare and with ap-
propriate brightness and contrast settings. 

 

Close the lid after placing or removing the microplates. 

 

Do not open the lid during measurements. 

 
BIOLOGICAL SAFETY 

 

It is the user’s responsibility to carry out appropriate  decontamination if hazardous 
material is spilt on or penetrates into the equipment. 

Summary of Contents for HiPo MPP-96

Page 1: ......

Page 2: ...2...

Page 3: ...tents 1 Safety Precautions 4 2 General Information 5 3 Getting started 6 4 Operation 7 5 Unit verification 10 6 Specifications 12 7 Maintenance 14 8 Warranty and Claims 14 9 EU Declaration of Conformi...

Page 4: ...power switch and external power supply are easily accessible during use Disconnect the unit from the external power supply before moving This equipment is connected and controlled by PC Please ensure...

Page 5: ...BestFit function for the selection of the best calibration curve Multiplex analysis up to 7 different tests on the same plate Qualitative assays the ability to install up to 8 types of controls weak p...

Page 6: ...pce Operating manual 1 copy Certificate of verification 1 copy Optional accessories Light filters custom 400 700 nm on request Verification microplate set on request Verification microplate set 3 3 Se...

Page 7: ...A1 is placed next to the A1 label tag fig 3 1 Close the lid Note Measurements can be affected by uneven or heterogeneous well contents Visually inspect the plate for foaming opaqueness bubbles or part...

Page 8: ...seconds 4 4 4 Press Start button The Measurement LED light fig 1 2 turns on Attention Do not open the lid during measurements 4 5 Measurement processing After an 8 second measurement the software aut...

Page 9: ...mat press XLS 96 well button 4 9 After completing the measurements open the lid and remove the microplate then close the lid Attention Do not leave the microplate in the unit 4 10 After finishing the...

Page 10: ...the certified verification plate Verification certificate is enclosed in the standard set and is valid one year from verification date All verification procedures must be performed only by qualified a...

Page 11: ...channels to test fig 5 4 and the number of measurements to perform fig 5 2 We recommend a minimum of 20 measure ments Press the Start button fig 5 3 5 2 4 When measurements are finished press the Sho...

Page 12: ...channel 1 6 9 Optical density measurement range 0 4 3 OD 6 10 Resolution 0 001 OD 6 11 Accuracy at 405 450 492 620 nm 0 000 2 000 OD 0 5 0 010 OD 2 000 3 000 OD 1 0 010 OD 6 12 Precision reproducibili...

Page 13: ...imum spectral power are not equal Table 1 96 well plate dimensions 96 well plate Dimension mm Plate height 14 35 Plate length 85 48 Plate width 127 76 First well position X 14 38 First well location Y...

Page 14: ...ments of Specifications provided the Customer follows the operation storage and transporta tion instructions 8 2 The warranted service life of the unit from the date of its delivery to the Customer is...

Page 15: ...ctive 2014 30 EU LVD Directive 2014 35 EU RoHS2 2011 65 EU WEEE 2012 19 EU Applicable Standards LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Gen...

Page 16: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 1 07 January 2017...

Reviews: