Page 15
V1.6.3
4.3.11. Push
“Enter” to confirm. Instrument will start to run the protocol program automatically until
whole processes are completed.
Note.
It will take times to complete the extraction according to reagent types.
4.3.12. At the end of the run, the instrument beeps briefly and the
LCM shows “Protocol Completed”.
4.3.13. Open the drawer.
4.3.14. Remove the elute tubes containing the purified nucleic acid.
Note.
Store the purified nucleic acids at 4
℃
for short-term storage or store at -70
℃
for
long-term storage.
4.3.15. Discard the used cartridges, all plastic consumables into biohazard waste. Do not reuse the
cartridges.
4.3.16. If you are not using the instrument, place the Sample Rack back to workplace, close the instru-
ment door and push
“Start” button for 2 seconds to get into “sleeping mode”. In addition, for
longer time not using the instrument turn the power switch off.
4.4.
After Extraction Operation
4.4.1. Take out the Elute Tubes (The extraction yields in them).
Note.
You can apply quality checking or do downstream study or storage them as what
you expect.
4.4.2. Remove the Cartridges and plastic disposables from the instrument and then dispose of them.
Summary of Contents for BioMagPure 12 Plus