Summary of Contents for BioMagPure 12 Plus

Page 1: ...Edition v1 6 3 ...

Page 2: ...www biosan lv Page 2 V1 6 3 ...

Page 3: ...n 5 3 Installation 7 4 Getting started 12 5 Cleaning and maintainance 16 6 Troubleshooting 17 7 Specifications 19 8 Replacement Optional Parts 19 9 Product Guarantees 20 10 About services 20 11 Notes for Transporting Shipping 20 12 Instrument transportation insert before transport 21 ...

Page 4: ... settings and maintenance This guide contains important information regarding the safe use of the BioMagPure 12 Plus system Please read this manual carefully before you start to run the system at the first time especially for Safety Information If there is any question about how to install or operate it please contact our certified distributors agents or email to our technical support centre suppo...

Page 5: ...wan US and Canada use a power cable that meets your country s standard or contact your local distributor Do not use the BioMagPure 12 Plus system with voltage other than the voltage specified on the device Do not use the BioMagPure 12 Plus system with a damaged power plug or a loose socket If there is dust on the prongs of the power plug or on the plug socket remove it with a dry cloth When you di...

Page 6: ...d be performed in an appropriate laboratory or workplace Note The kits are not supplied with the BioMagPure 12 Plus system Select the desired kit s and order it them separately 2 3 About Samples Warning Biohazard Always wear appropriate gloves a mask and safety goggles etc when handling any infectious samples 2 4 About Infectious Wastes When handling or disposing of infectious materials follow the...

Page 7: ...rt s A B C D E F A BioMagPure Instrument 1 B Power Cable 1 C Sample Tube Rack 2 D Barcode Reader 1 E User s Manual 1 F M axis Module 1 3 2 Please make all the components are free from damages as soon as you get the system If any damage found please contact your local representative for the instant support Note Biosan s global warranty does not cover the damages from the transportation or improper ...

Page 8: ...to 40 C and humidity is 30 to 80 RH non condensing A location that is flat and stable with no vibration A location away from direct sunlight Block the sunlight by closing curtains or blinds as neces sary A location which is well ventilated and not dusty A location where the temperature does not go up and down suddenly Warming a cold room suddenly or moving the BioMagPure 12 Plus system from a room...

Page 9: ...ing operation 30 80 During down time 10 80 Max wet bulb temperature C During operation 29 non condensing During down time 29 non condensing Temperature gradient C hr 12 or less non condensing Humidity gradient RH day 30 or less non condensing Altitude operating m 1600 or less 3 6 Unpacking of the BioMagPure 12 Plus system 3 6 1 Open the packing box and take out the instrument and related accessori...

Page 10: ...at least 3 turns 3 6 5 How to remove the transportation insert inside the instrument following the steps below 3 6 5 1 Plug in the power cable to the instrument and connect to electric outlet 3 6 5 2 Turn the power switch on 3 6 5 3 Press START button The piston module will going up and you can release the transportation insert 3 7 Overall View Front view with Door closed Front view with Door open...

Page 11: ... 3 LCM Panel and Control Pad Control pad close view LCM displays information of operation instruction guide and program process status Control pad is used for selection of functions and input of program and test codes Rear panel ...

Page 12: ...dividual protection such as gloves masks and safety gog gles For the preparation of samples refer to the handbook of each kit Following items are provided with reagent kits reagent cartridge reaction chamber tip holder small tip optional filtered tip sample tube elute tube piercing pin 4 3 Extraction 4 3 1 Turn the power switch on and waiting for the LCM screen turn on and shows BioMagPure 12 Plus...

Page 13: ...ed tip are displayed below leave middle row empty 4 3 5 Load one Reagent Cartridge and one set of plastic disposable per sample Set Cartridges in the order of the number from left to right Make sure that Cartridges are inserted in to the Cartridge Tray tightly You can load 1 12 cartridges on the tray depending on the number of samples that you wish to process ...

Page 14: ...ure the caps of Elute Tube are open as the figure shown above Use two hands to handle the Sample Tray Make sure the Sample Tray be placed correctly in the instrument 4 3 8 Close the Drawer 4 3 9 Scan the protocol barcodes to select purification protocol sample volume and elute volume Note There is one protocol barcode paper enclosed in the reagent kit box Protocol s name sample volume and elution ...

Page 15: ...rage or store at 70 for long term storage 4 3 15 Discard the used cartridges all plastic consumables into biohazard waste Do not reuse the cartridges 4 3 16 If you are not using the instrument place the Sample Rack back to workplace close the instru ment door and push Start button for 2 seconds to get into sleeping mode In addition for longer time not using the instrument turn the power switch off...

Page 16: ...side Preventive Service Engineer Annually 5 2 Clean the Sample Tray with mild detergent and rinse with deionized water Allow the parts to dry before use Clean and disinfect the platform surface by wiping with deionized water followed by 75 ethanol 5 3 Clean the instrument body by removing dust gently with a dry soft cloth If the outside of the BioMagPure 12 Plus system is heavily soiled or if any ...

Page 17: ...rs Turn off the power and remove all samples and plastic wares Contact your local representatives Bubbles formed during extraction Missed adding sample or sample vol ume is lower than the recommended volume Be sure to add the sample to tubes prior to starting the protocol To ensure proper mixing of reagents in the tip and prevent bubble formation during mixing make sure the sample volume is at lea...

Page 18: ...to the BioMagPure purification No DNA recovered Magnetic beads stored or handled improperly Store cartridge containing the beads at room tempera ture Do not freeze the cartridge as the beads may be irrepa rably damaged Make sure that the beads are in solution at all times and do not dry Dried beads are non functional Elute containing DNA is discoloured Magnetic beads present in the elute Remove an...

Page 19: ...2 samples per batch Sample volume handling 100 1200μL Processing time See purification kit manual for details Heat block temperature 60 C to 70 C assuming the room temperature of 25 C Protocol input Barcode Reader UV light 30 minutes 250 nm Display LCD Screen Instrument dimensions 56 cm W x 51 cm D x 59 cm H Weight 60 kg Input power AC 100 240 V 240 VA 50 60 Hz Fuse F3 15A 250V 7 1 Applicable stan...

Page 20: ...facturer 9 3 2 Any malfunctions or damages during transportation or due to dropping of the product by a user 10 About services 10 1 Before you ask for device repairs refer to 4 Troubleshooting to check the problems If the problems cannot be solved contact your local representative or the relative customer support 11 Notes for Transporting Shipping 11 1 Before carrying the BioMagPure 12 Plus system...

Page 21: ...re transport 12 1 Press 0 to enter the menu 12 2 Press 2 for engineering menu 12 3 Press 3 to choose FIXING 12 4 Open the front door of the unit Put the transportation insert on the cartridge plate and close the door 12 5 Lock the X axis with 4 mm hex tool as shown below ...

Page 22: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv ...