6.  Maintenance 

6.1. 

If the unit requires maintenance, disconnect the unit from the electric circuit 
and contact Biosan or your local Biosan representative. 

6.2. 

All  maintenance  and  repair  operations  must  be  performed  only  by  qualified 
and specially trained personnel. 

6.3. 

Standard ethanol (75%) or other cleaning agents recommended for cleaning 
of  laboratory  equipment  can  be  used  for  cleaning  and  decontamination  of 
the unit. 

6.4. 

Fuse replacement.

 Disconnect from the power supply socket. Remove the 

power  cable  from  the  rear  of  the  unit.  Pull  out  the  fuse  drawer  by  applying 
leverage in recess (A). Remove the fuse from the holder.  Check and replace 
with the correct fuse if necessary, 230V 

– M 630mA (type 

M

 - time lag: 

M

e-

dium). 

Summary of Contents for Automatic Luminescence Analyser ALA1/4

Page 1: ......

Page 2: ...2...

Page 3: ...3 Contents 1 Safety Precautions 4 2 General Information 5 3 Getting started 6 4 Operation 7 5 Specifications 8 6 Maintenance 9 7 Warranty and Claims 10...

Page 4: ...use an ungrounded extension lead Disconnect the unit from electric circuit before moving Switch the unit off and disconnect the power cord plug from power socket to disconnect the unit from electric...

Page 5: ...on The close tubes method ensures a biosafety conditions in the DNA diagnostic Labs The product consists of the flowing main parts A rotor with tubes sockets fixed on the motor axis Optical module Mic...

Page 6: ...r 0 5 ml tubes 1 pce Rotor for 0 2 ml tubes 1 pce Data Cable 1 pce Software installation disc 1 pce Software installation manual 1 copy Software operating manual 1 copy Software Quick Start 1 copy Ope...

Page 7: ...ation Install the rotor into the analyser The positioning pin on the base inside the unit must match into the opening in the rotor Secure the new rotor carefully and turning the latch tightly Caution...

Page 8: ...length 570 nm emission wavelength 610 nm 5 6 Fourth Channel CY 5 excitation wavelength 630 nm emission wavelength 670 nm 5 7 Rotor capacity R 036 36 x 0 5 ml tubes R 048 48 x 0 2 ml tubes 5 8 Measurem...

Page 9: ...personnel 6 3 Standard ethanol 75 or other cleaning agents recommended for cleaning of laboratory equipment can be used for cleaning and decontamination of the unit 6 4 Fuse replacement Disconnect fro...

Page 10: ...ck availability of extended warranty 7 3 Warranty covers only the units transported in the original package 7 4 If any manufacturing defects are discovered by the Customer an unsatisfac tory equipment...

Page 11: ...11...

Page 12: ...SIA Biosan Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 6742 6137 Fax 371 6742 8101 http www biosan lv Edition 2 01 October 2015...

Reviews: