Noninvasive Blood Pressure Monitoring System 

After setting the Date/Time Properties, click “OK” at the top right of the window to accept the new settings. To 

discard settings without saving changes, click the “X” button. 
The “Pulse gain” and the “Cut Off Pressure” are adjusted by moving the sliders left or right. Click OK to accept 

any modified settings. Click Cancel to discard. 

2.5.2. Options 

Use the Options Tab (see Figure 9) to adjust the triggering parameters. If an external trigger to be used to initiate 

the start of a measurement, enable the “Trigger Input active” option. 

 

Figure 9: Options Tab Settings 

If “Auto Save after trigger” is enabled, the device will automatically save the last measurement. 

2.5.3. Calibration 

In the Calibration Tab (see Figure 10)  the pressure sensor calibration can be checked and recalibrated if 

necessary.  

 

Figure 10: Calibration Tab Settings 

The Calibration Tab displays the raw sensor values for two calibration points at 0 mmHg and 100 mmHg, current 

pressure and raw signal values, and the calibration line formula.  
The “Cal 1” and “Cal 2” buttons are used for measuring and setting the raw signal values for each calibration 

point. Pressing the “Restore default calibration” button reverts the calibration setting to it’s factory default. 

2.6. Results List 

The Results List window (see Figure 11) lists the previously saved measurements. For each measurement, the 

Result Number, Systolic, Diastolic, Mean pressures, heart rate, description and save date/time are displayed. 
Measurements that do not return any results are identified by a value of -1.0.  

[email protected]

 

Page 9 of 18 

Summary of Contents for NIBP250

Page 1: ...NIBP250 User s Guide Noninvasive Blood Pressure Measurement System 42 Aero Camino Goleta CA 93117 Telephone 805 685 0066 Facsimile 805 685 0067 support biopac com www biopac com...

Page 2: ...rs 13 4 Using the NIBP250 with a BIOPAC MP Acquisition System 14 4 1 Connections 14 4 2 Software Setup in AcqKnowledge 14 5 Technical Specifications 15 6 Troubleshooting 16 6 1 Tail Pulse signals are...

Page 3: ...low diastolic pressure When blood flow is nil cuff pressure exceeding systolic pressure or unimpeded cuff pressure below diastolic pressure cuff pressure will be essentially constant When blood flow i...

Page 4: ...to reach the selected temperature 7 Remove the RESTRAINER from the Animal Heating Chamber 8 Connect the Tail Cuff Sensor to the animal s tail 9 Verify that the sensor fits the tail correctly The sens...

Page 5: ...7 USB Memory Slot 3 Pulse Signal Analog Output 8 SD Card Slot 4 Pressure Signal Analog Output 9 Tail Sensor Input 5 Trigger Input 10 Tail Cuff Tubing Port 2 3 Sensor Figure 3 shows Tail Cuff Sensor an...

Page 6: ...ttons Button Functionality Button Functionality Moves the cursor to the left or right by one step If Peak by peak is enabled the cursor moves to the previous or next peak Saves the last measurement Se...

Page 7: ...option is selected the value of the peak is displayed If unchecked the value of the measurement sample is displayed To record systolic and diastolic points move the cursor to the desired axis position...

Page 8: ...6 Menu window 2 5 1 General Tab The General Tab see Figure 7 is where the NIBP200 date and time is set or adjusted Press the Set button to display the Date Time Properties see Figure 8 Figure 7 Genera...

Page 9: ...2 5 3 Calibration In the Calibration Tab see Figure 10 the pressure sensor calibration can be checked and recalibrated if necessary Figure 10 Calibration Tab Settings The Calibration Tab displays the...

Page 10: ...ns the leakage valve to release the pressure After the pressure is fully released the acquisition stops 3 3 Analysis Following the acquisition the unit automatically detects the pulse peaks after whic...

Page 11: ...the pulse with the highest peak and then press the Diastolic button The system will record and display the diastolic blood pressure value You may change your cursor peak positions at anytime during t...

Page 12: ...pressure value Heart rate value Description text Data sample period in seconds Number of data lines Pressure value Pulse value at time 0 s Pressure value Pulse value at time 0 004s Pressure value Pul...

Page 13: ...pressure value shown by the manometer and the Current Pressure value on the NIBP250 s screen 9 If the two values match the calibration is validated If not recalibration is necessary 3 9 2 Recalibrati...

Page 14: ...with BIOPAC MP36 35 Data Acquisition Unit with BSL PRO software use the CBL35 PRE for Step 1 and CBL35 PLS for Step 2 NOTE BIOPAC recommends connecting the CBL150 PRE cable to the first channel A1 so...

Page 15: ...e Sensitivity 0 1 mmHg Pressure Signal Output 300 mmHg 3 Volt DC Pulse Gain Levels x1 x2 x4 x5 x8 x16 x32 adjustable Pulse Signal Output 0 5 Volt DC LCD Display 7 800x480 TFT User Interface Resistive...

Page 16: ...sor in a different location on the tail The optimal location is between the mid point of tail and base of tail spinal column 6 2 Compressor runs continuously Immediately Turn off the NIBP250 system Re...

Page 17: ...1 mm RXTCUFSENSOR13 Tail Cuff Sensor 13 mm RXCUFSEN9 5 MRI Tail Cuff Sensor for MRI 9 5 mm RXCUFSEN11 MRI Tail Cuff Sensor for MRI 11 mm RXCUFSEN13 MRI Tail Cuff Sensor for MRI 13 mm RXRESTRAINER S Re...

Page 18: ...Representative Provide the following information Your Name Unit Serial Number Your Institution Sensor Size Address Describe the problem Telephone Number Facsimile Number E mail Address You will be ad...

Reviews: