NIBP250 User’s Guide 

4. Using the NIBP250 with a BIOPAC MP Acquisition System 

4.1. Connections 

1.  Connect the CBL150-PRE cable* 

a.  BNC to the pressure output on the back panel of the NIBP250. 
b.  Other end to A1 on the front of the UIM100C unit 

2.  Connect the CBL150-PLS cable* 

a.  BNC to the pulse output on the back panel of the NIBP250. 
b.  Other end to A2 on the front of the UIM100C unit. 

3.  Connect the IRSENSOR 

a.  Black cord to the sensor input on the back panel of the NIBP250. 
b.  Tubing in the cuff  on the front panel of the NIBP250. 

4.  Connect the power 

a.  AC300 adapter to the 12V DC input on the back panel of the NIBP250. 
b.  AC300 to Mains power. 

5.  Switch the NIBP250 POWER on. 

*CBL150-PRE and CBL150-PLS cables are used with BIOPAC MP150 Data Acquisition Unit with Acq

Knowledge 

software only. For use with BIOPAC MP36/35 Data Acquisition Unit with BSL 

PRO

 software, use the CBL35-PRE 

for Step 1 and CBL35-PLS for Step 2. 

NOTE: 

BIOPAC recommends connecting the CBL150-PRE cable to the first channel (A1) so that the pressure 

wave is automatically displayed above the pulse waveform. 

4.2. Software Setup in Acq

Knowledge 

1. Laun ch Acq

Knowledge

 

2.  Select the “Create/Record a new experiment” option. 
3.  Select MP150 > Set Up Channels > “Add New Module...” 

a.  From the new module list, select UIM100C-A1 (or whichever channel CBL150-PRE pressure 

cable is connected to) and click “Add.” 

b.  From the UIM100C Transducer list, select “NIBP250 – Small Animal Tail BP, Pressure” and click 

OK. 

c. Click  “Calibrate” in the resulting NIBP250 Calibration dialog. 

 

4.  Repeat  “Add New Module…” portion of Step 3. 

a.  From the new module list, select UIM100C-A2 (or whichever channel CBL150-PLS pulse cable is 

connected to) and click “Add.” 

b.  From the UIM100C Transducer list, select “NIBP250 – Small Animal Tail BP, Pulse” and click OK. 

 
 

Page 14 of 18 

www.biopac.com 

Summary of Contents for NIBP250

Page 1: ...NIBP250 User s Guide Noninvasive Blood Pressure Measurement System 42 Aero Camino Goleta CA 93117 Telephone 805 685 0066 Facsimile 805 685 0067 support biopac com www biopac com...

Page 2: ...rs 13 4 Using the NIBP250 with a BIOPAC MP Acquisition System 14 4 1 Connections 14 4 2 Software Setup in AcqKnowledge 14 5 Technical Specifications 15 6 Troubleshooting 16 6 1 Tail Pulse signals are...

Page 3: ...low diastolic pressure When blood flow is nil cuff pressure exceeding systolic pressure or unimpeded cuff pressure below diastolic pressure cuff pressure will be essentially constant When blood flow i...

Page 4: ...to reach the selected temperature 7 Remove the RESTRAINER from the Animal Heating Chamber 8 Connect the Tail Cuff Sensor to the animal s tail 9 Verify that the sensor fits the tail correctly The sens...

Page 5: ...7 USB Memory Slot 3 Pulse Signal Analog Output 8 SD Card Slot 4 Pressure Signal Analog Output 9 Tail Sensor Input 5 Trigger Input 10 Tail Cuff Tubing Port 2 3 Sensor Figure 3 shows Tail Cuff Sensor an...

Page 6: ...ttons Button Functionality Button Functionality Moves the cursor to the left or right by one step If Peak by peak is enabled the cursor moves to the previous or next peak Saves the last measurement Se...

Page 7: ...option is selected the value of the peak is displayed If unchecked the value of the measurement sample is displayed To record systolic and diastolic points move the cursor to the desired axis position...

Page 8: ...6 Menu window 2 5 1 General Tab The General Tab see Figure 7 is where the NIBP200 date and time is set or adjusted Press the Set button to display the Date Time Properties see Figure 8 Figure 7 Genera...

Page 9: ...2 5 3 Calibration In the Calibration Tab see Figure 10 the pressure sensor calibration can be checked and recalibrated if necessary Figure 10 Calibration Tab Settings The Calibration Tab displays the...

Page 10: ...ns the leakage valve to release the pressure After the pressure is fully released the acquisition stops 3 3 Analysis Following the acquisition the unit automatically detects the pulse peaks after whic...

Page 11: ...the pulse with the highest peak and then press the Diastolic button The system will record and display the diastolic blood pressure value You may change your cursor peak positions at anytime during t...

Page 12: ...pressure value Heart rate value Description text Data sample period in seconds Number of data lines Pressure value Pulse value at time 0 s Pressure value Pulse value at time 0 004s Pressure value Pul...

Page 13: ...pressure value shown by the manometer and the Current Pressure value on the NIBP250 s screen 9 If the two values match the calibration is validated If not recalibration is necessary 3 9 2 Recalibrati...

Page 14: ...with BIOPAC MP36 35 Data Acquisition Unit with BSL PRO software use the CBL35 PRE for Step 1 and CBL35 PLS for Step 2 NOTE BIOPAC recommends connecting the CBL150 PRE cable to the first channel A1 so...

Page 15: ...e Sensitivity 0 1 mmHg Pressure Signal Output 300 mmHg 3 Volt DC Pulse Gain Levels x1 x2 x4 x5 x8 x16 x32 adjustable Pulse Signal Output 0 5 Volt DC LCD Display 7 800x480 TFT User Interface Resistive...

Page 16: ...sor in a different location on the tail The optimal location is between the mid point of tail and base of tail spinal column 6 2 Compressor runs continuously Immediately Turn off the NIBP250 system Re...

Page 17: ...1 mm RXTCUFSENSOR13 Tail Cuff Sensor 13 mm RXCUFSEN9 5 MRI Tail Cuff Sensor for MRI 9 5 mm RXCUFSEN11 MRI Tail Cuff Sensor for MRI 11 mm RXCUFSEN13 MRI Tail Cuff Sensor for MRI 13 mm RXRESTRAINER S Re...

Page 18: ...Representative Provide the following information Your Name Unit Serial Number Your Institution Sensor Size Address Describe the problem Telephone Number Facsimile Number E mail Address You will be ad...

Reviews: