- Avoid contact with dripping or splashing liquids.
- The minimum blood sample size to test using the RIGHTEST Blood Glucose Monitoring System
GM280B is 0.75
μL
: ( )
5
4
Caution
Caution
Blood sample size above 3.0
μL
might contaminate the test strip port and the meter.
Blood sample size below 0.75
μL
may cause an inaccurate result or may prevent a meter reading.
An Er4 reading will be displayed if the sample size is too small. In this case, repeat the test with a
new test strip.
Sample Size Example
Important Safety Notes:
- All parts of the kit are considered biohazardous and can potentially transmit infectious
diseases, even after following the cleaning and disinfecting procedures.
- Users should wash their hands thoroughly with soap and water before and after handling
the meter, lancing device, or test strips.
0.75
μL
1.0
μL
1.5
μL
2.0
μL
3.0
μL
- Before using the RIGHTEST Blood Glucose Monitoring System GM280B to test your blood
glucose, please read all of the instructions and conduct all of the tests including the quality control
test.
- Please perform the quality control test regularly to make sure the test results are accurate.
- The RIGHTEST Blood Glucose Meter GM280B can only be used with RIGHTEST Blood Glucose
Test Strips GS280 and RIGHTEST Control Solution GC550. The application of other Test Strips or
control solutions can lead to incorrect results.
- RIGHTEST Blood Glucose Monitoring System GM280B is intended for self- testing outside the
body (
in vitro
diagnostic use), it is not use for diagnosis of, or screening for diabetes.
- RIGHTEST Blood Glucose Monitoring System GM280B is NOT use for critically ill patients and
neonate.
- If RIGHTEST Meter and Test Strips are exposed to temperature environments out of range for the
meter – below 10 °C (50 °F) or above 40 °C (104 °F) – please wait 30 minutes before testing again.
- Follow all environmental protection regulations when disposing of batteries, strips and lancets.