Understanding Control Test Results

Recalling Test Results

The RIGHTEST Blood Glucose Meter

B is able to automatically store a maximum of 500 test 

 GM280

results with time and date. If your meter has stored 500 results, the newest test result will replace 
the oldest one.

To recall your test memory, start the meter without a test strip inserted.

1. Press the Main button to switch from the Time Mode to Memory Mode. You 

will see the “mem” icon in the upper left corner of the display. The latest test 
result will appear with the number 

"

 1

 "

.

2. Use either the Left or Right button on the side to review all previous results 

with date and time. Results display from the most recent (Sequence no. 

"

 1 

"

to the oldest (Sequence no. 

"

 500 

"

) in the lower right hand corner of the 

screen.

3. The Quality Control Test result can be recalled from the saved data. When 

"            " appears next to the data, the test was performed using the  
RIGHTEST Control Solution GC550. The control solution test result will not be 
used when recalling test average results.

Your control solution test results should fall within the control solution range. If the results are within 
the range, the RIGHTEST Blood Glucose Monitoring System GM280B is working correctly.

Possible reasons your Control Solution results are out of the range :
- Your RIGHTEST Control solution GC550 is expired or was first opened more than 3 months ago.
- Your RIGHTEST Blood Glucose Test Strip GS280 has expired.

  

 

- You left the cap of the  Blood Glucose Test Strip vial or the  control solution off for a period of time.
- You did not perform the test procedure correctly. 

- The

 RIGHTEST 

B Blood Glucose Meter or

 RIGHTEST 

Blood Glucose Test Strip GS280 

GM280

have malfunctioned. 

   
If RIGHTEST Control Solution GC550 results are out of range, your RIGHTEST Blood Glucose 
Monitoring System GM280B may not be working properly. Repeat the Quality Control Test. If your 
control solution results are still out of range, do not use RIGHTEST Blood Glucose Meter GM280B 
to test your blood glucose. Please contact customer service.

47

46

Summary of Contents for RIGHTEST GM280B

Page 1: ...ATION No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan Tel 886 4 2369 2388 Fax 886 4 2261 7586 http www bionime com E mail info bionime com Made in Taiwan BIONIME GmbH Tramstrasse 16 9442...

Page 2: ......

Page 3: ...y questions or concerns please contact your local Bionime Customer Service or email to rightest bionime com RIGHTEST Blood Glucose Monitoring System GM280B is intended to be used for the quantitative...

Page 4: ...RIGHTEST Blood Glucose Monitoring System GM280B to test your blood glucose please read all of the instructions and conduct all of the tests including the quality control test Please perform the qualit...

Page 5: ...ts under dehydrated hypo osmolar or in shock Alternative site testing AST should only be performed during steady state times when glucose is not changing rapidly Limitations Please refer to the Altern...

Page 6: ...he RIGHTEST Blood Glucose Monitoring System GM280B The RIGHTEST Blood Glucose Monitoring System GM280B The RIGHTEST Blood Glucose Meter GM280B The RIGHTEST Blood Glucose Test Strip GS280 Before Testin...

Page 7: ...ency Card 2 RIGHTEST Blood Glucose Meter GM280B with 1 CR2032 battery installed 3 RIGHTEST Blood Glucose Test strips GS280 10 or 25 pcs with Package Insert 4 RIGHTEST Control Solution GC550 with Packa...

Page 8: ...GHTEST Blood Glucose Meter GM280B Right Button Battery Cover Meter Battery The RIGHTEST Blood Glucose Meter GM280B Left Button Main Button Display Window Test Strip Port 2D barcode with serial number...

Page 9: ...p into meter Appears when an error occurs Current date under time mode or testing date under memory mode Warns when the operational temperature limit is exceeded during testing Indicates the time in 1...

Page 10: ...F and dry location 90 relative humidity Do not expose to direct sunlight or heat For detailed information please refer to the RIGHTEST Test Strip GS280 insert If the RIGHTEST GM280B meters and strips...

Page 11: ...se follow the local regulation to properly discard a used battery 4 The RIGHTEST Blood Glucose Meter will enter Self Testing Mode automatically when GM280B replacing the battery all symbols will appea...

Page 12: ...n for 7 seconds During this time the screen will go blank until you hear a beep After the beep the meter will turn on into the Setting Mode The display screen will show setting data Setting Up Your Me...

Page 13: ...o change any settings during Meter Set up for over 2 minutes the RIGHTEST Blood Glucose Meter will leave setting mode and power off automatically GM280B Setting Up Your Meter Setting the Date Time and...

Page 14: ...M280B buttons are pressed or no strip is inserted Turning On Off the Meter Handling the RIGHTEST Blood Glucose Test Strip GS280 How to handle the RIGHTEST Blood Glucose Test Strip GS280 Inserting the...

Page 15: ...Test Strip GS280 out of the test strip port Please follow your local regulations and discard used strips properly 1 3 Getting Ready for Testing Before performing a blood glucose test prepare the item...

Page 16: ...age skin for thick or calloused skin 1 2 3 4 5 6 Performing a Blood Glucose Test 7 Hold the base in one hand and pull on the plunger with the other hand The device will be cocked Release the plunger a...

Page 17: ...rt If the blood sample was insufficient discard the test strip and repeat with a new RIGHTEST Blood Glucose Test Strip GS280 Performing a Blood Glucose Test 15 Countdown Mode will begin on the display...

Page 18: ...rts of this kit are considered biohazards and can potentially transmit infectious diseases even after you have performed the cleaning and disinfecting procedure Users should clean the meter and wash h...

Page 19: ...e screen CAUTION No suggest to test blood sugar when the flight is strictly restrict the Bluetooth connection Please turn off you meter to avoid the auto on of Bluetooth RIGHTEST CARE App is designed...

Page 20: ...monitors CGMs Use the clear cap provided with RIGHTEST Lancing Device when testing sites other than fingertip The normal lancing device cap for your palm or forearm the blood sample may not be suffici...

Page 21: ...sed test strips and lancets Make sure your blood sample covers the whole area of the view window to get an accurate test result An insufficient blood sample will result in an error message Er4 If this...

Page 22: ...ur test strip vial label NOTE If you want purchase control solutions please contact your authorized Bionime representative or Bionime customer service printed on manual cover The RIGHTEST Meter B disp...

Page 23: ...Quality Control Test Be Performed To ensure that your RIGHTEST Blood Glucose Meter B and RIGHTEST Blood Glucose Test GM280 Strip are working properly GS280 To confirm that you are following the corre...

Page 24: ...e bottle of control solution and may cause the degeneration of the control solution This could contaminate the meter via the test strip port Keep the test strip port clean and dry Clean immediately if...

Page 25: ...Solution GC550 The control solution test result will not be used when recalling test average results Your control solution test results should fall within the control solution range If the results are...

Page 26: ...r 90 days averages 3 The number shown in the lower right hand corner indicates how many test results have been calculated The RIGHTEST Blood Glucose Meter B provides several average test results View...

Page 27: ...se with the meter and lancing device Maintaining Keep your meter and test strip free of dust water or any other liquid Store the meter in the carrying case when not in use If your meter is dropped or...

Page 28: ...meters and strips are exposed to temperature environments out of range for the meter 10 C 50 F or above 40 C 104 F or a high temperature difference please wait 30 minutes before measurement CAUTION Us...

Page 29: ...he meter should work normally after finishing the above steps If the meter still does not work please contact your local customer service Bionime Strip Error 1 When the test strip is inserted incorrec...

Page 30: ...0 104 F 10 90 One CR2032 battery 3 Voltage Approximately 1000 standard tests Turns off automatically 2 minutes after last user action To turn off manually press the button for 3 seconds Specification...

Page 31: ...of the RIGHTEST Blood Glucose Monitoring System GM280B are considered biohazardous and can potentially transmit infectious diseases Please follow your local regulations to properly dispose of the use...

Page 32: ...Medical Device Directive 93 42 EEC EC Rep Bionime GmbH Tramstrasse 16 9442 Berneck Switzerland Disposable Sterile Lancets Manufacture information is as package printed Expected Glucose Values Referenc...

Reviews: