Maintain the Products
The cleaning procedure is to remove dust, blood and body fluid from the surface and should be
performed whenever the meter or lancing device is visibly dirty. Performing the cleaning procedure
once per week is recommended. The disinfecting procedure is necessary to kill pathogens such as
HIV, HBV and HCV on the device. This procedure should be performed regularly; once per week is
recommended.
If the meter is being operated by a second person who is providing testing assistance to the user,
the meter and lancing device should be cleaned and disinfected prior to use by the second person.
CAVIWIPES DISINFECTING TOWELETTES, manufacturer: Metrex. It is with Isopropanol as the
active ingredient, have been shown to be safe for use with the meter and lancing device.
Maintaining
Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying
case when not in use. If your meter is dropped or damaged, perform a quality control test with the
control solution before performing a blood glucose test.
Maintain the Products
2
1
3
2 min
NOTE
- Clean and disinfect the outside of the device only. Do not remove battery cover when
cleaning and disinfecting.
Cleaning and Disinfecting frequency: at least once a week.
To clean the meter:
1. Thoroughly wipe the entire surface of the meter with disinfecting
wipes listed above to clean any possible dirt, dust, blood and other
body fluids.
To disinfect the meter:
2. Take another disinfecting wipe and wipe the meter thoroughly (Note:
All blood and body fluids should be cleaned from surface before
performing the disinfecting procedure).
3. Allow the surface to remain wet for 2 minute.
4. Allow to air dry.