V 1.31
2021. 05
163
/
171
Doc. # : BN-OP-C7
Immunity to Conducted
Disturbances Induced by
RF fields
IEC 61000-4-6
3 V
0.15 MHz - 80 MHz
6 V in ISM bands
Between 0.15 MHz and 80 MHz
80% AM at 1 kHz
Power Frequency Magnetic
Field Immunity
IEC 61000-4-8
30 A/m
50 Hz and 60 Hz
Voltage dips
IEC 61000-4-11
0 %
U
T
: 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and
315°
0 %
U
T
; 1 cycle
and
70 %
U
T
; 25/30
cycles
Single phase: at 0º
Voltage interruptions
IEC 61000-4-11
0 %
U
T
; 250/300 cycle
Electromagnetic compatibility_ Guidance and manufacturer's declaration
Guidance and manufacturer’s declaration – electromagnetic emissions
The Cardio7e/Cardio7 is intended for use in the electromagnetic environment specified below.
The customer or the user of the Cardio7e/Cardio7 should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Cardio7e/Cardio7 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The Cardio7e/Cardio7 is suitable for use in all
establishments other than domestic, and may be
used in domestic establishments and those
directly connected to the public low-voltage power
Harmonic emissions
IEC 61000-3-2
Class A
Summary of Contents for Cardio7
Page 21: ...V 1 31 2021 05 21 171 Doc BN OP C7 Front View Printer Cover Switch...
Page 61: ...V 1 31 2021 05 61 171 Doc BN OP C7 Rhythm Report Form 3CH Rhythm Report Form 6CH...
Page 62: ...V 1 31 2021 05 62 171 Doc BN OP C7 Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Page 63: ...V 1 31 2021 05 63 171 Doc BN OP C7 Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Page 67: ...V 1 31 2021 05 67 171 Doc BN OP C7 Record Report Form BEAT REPORT ST Map Bar...