V 1.31
2021. 05
122
/
171
Doc. # : BN-OP-C7
When selecting the
‘
Export
’
menu, a small menu window as below will appear and
it will disappear when selecting one of None, Server, USB.
If it is ‘None,’ ECG data will not be externally transferred; if is 'USB,' ECG data will be
automatically saved to the external USB memory. In addition, if it is set up as ‘
Server
,’ data
will be automatically transferred to PC Server IP among Network options. (Default: None)
Note
When you remove the external device after sending to USB Memory, you
should click icon
that is correct to each device
On top of the main screen and remove the device.
Otherwise data can be damaged.