1) Location of Electrodes
Attach electrodes to the patient’s body to record an
electrocardiogram of twelve
standard leads [ I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6], as shown below.
Limb electrodes are located as follows.
-
RL (N): Right Leg
-
LL (F): Left Leg
-
RA (R): Right Arm
-
LA (L): Left Arm
Chest electrodes are located as follows.
-
V1 (C1): Fourth Intercostal Space at The Right Border of The Sternum
-
V2 (C2): Fourth Intercostal Space at The Left Border of Sternum
-
V3 (C3): Midway Between Location V2 and V4
-
V4 (C4): At the Mid-Clavicular Line in The Fifth Intercostals Space
-
V5 (C5): At the Another Axillary Line on The Same Horizontal Level as V4
-
V6 (C6): At the Anterior Axillary Line on The Same Horizontal Level as V4 and V5