Rev. 1.1 5.ECG 89
BM3VET Touch User’s Manual
Attaching Electrodes to the Animal
1. Shave excess hair. With a piece of cotton pad moistened with alcohol, clean the Animal
’s skin
where the electrodes should be mounted. Avoid wrinkled or uneven skin areas. Wipe off the alcohol
with a dry cotton pad.
2. Open the electrode package and take out the electrode.
3. Remove the backing paper from the electrode. Be careful not to touch the adhesive side.
4. Attach the disposable electrode to the previously cleaned skin. Avoid wrinkled and uneven skin
areas.
5. The electrode lead is connected to the monitor onto the electrode.
6. Fasten the electrode lead to the skin with surgical tape with an extra length of wire between the
tape and the electrode. This prevents body movement from moving the electrode lead.
Note
To maintain good contact between the electrode and skin, check that the paste of the
disposable electrode is not dry.
When contact of the disposable electrode becomes poor, replace the electrode with a new
one immediately. Otherwise, contact impedance between the skin and electrode increase
and the correct ECG cannot be obtained.
If the contact is bad before the expiration date on the package, replace the electrode with a
new one.
To obtain a stable ECG wave form rub the skin with
“skin Pure” skin preparation gel or
tincture of Benzion.
Use only the CE certified disposable electrode.
Summary of Contents for BM3VET Touch
Page 1: ...BM3VET Touch User s Manual Veterinary Patient Monitor Rev 1 1 20160401...
Page 29: ...Rev 1 1 1 BASIC 29 BM3VET Touch User s Manual Features of Main Body FRONT BACK...
Page 88: ...Rev 1 1 5 ECG 88 BM3VET Touch User s Manual IEC 3LEAD AHA 3LEAD IEC 5LEAD AHA 5LEAD...
Page 91: ...Rev 1 1 5 ECG 91 BM3VET Touch User s Manual 5 Position of 5 Lead RL V LL LA RA...
Page 92: ...Rev 1 1 5 ECG 92 BM3VET Touch User s Manual Position of 3 Lead Wire Electrodes LL LA RA...