User Manual Smart Storage Agitator 30
Version
: 1.0
Validation date of the manual: 01/08/2021
21 of 29
(IEC61000-3-2)
Voltage variations, voltage
fluctuations and flicker
(IEC61000-3-3)
Compliant
Magnetic and electromagnetic immunity
The SST-A30 is intended to be used in the magnetic and electromagnetic environment
described in the table below. The user and the installer must guarantee the compliance of the
electromagnetic environment.
Immunity test
Test level according to
IEC60601
Compliance level
Electromagnetic
environment/comments
Electrostatic
discharge (ESD)
(IEC61000-4-2)
± 8 kV contact
discharge
± 2 kV; ± 4 kV; ± 8 kV;
± 15 kV air discharge
± 8 kV contact
discharge
± 15 kV air discharge
Professional healthcare
facility environment.
Fast transient
bursts
(IEC61000-4-4)
± 2 kV for power
supply lines
± 2 kV for power supply
lines
± 1 kV for signal ports
Professional healthcare
facility environment.
Surges
(IEC61000-4-5)
± 1 kV in Differential
mode
± 2 kV in common
mode
± 1 kV in Differential
mode
± 2 kV in common
mode
Professional healthcare
facility environment.
Power-frequency
magnetic field
(IEC61000-4-8)
30 A/m
30 A/m
Professional healthcare
facility environment.
Voltage dips, short
interruptions, and
voltage variations
(IEC61000-4-11)
0% UT for 0.5 cycles
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% UT for 1 cycle
70% UT
for 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single-phase: at 0°
0% UT for 0.5 cycles
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0% UT for 1 cycle
70% UT
for 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single-phase: at 0°
Professional healthcare
facility environment.
Voltage
interruptions
(IEC61000-4-11)
0 % UT;
for 250 cycles at 50 Hz
for 300 cycles at 60 Hz
0 % UT;
for 250 cycles at 50 Hz
for 300 cycles at 60 Hz
Professional healthcare
facility environment.