![biolog id SMART STORAGE AGITATOR 30 User Manual Download Page 10](http://html1.mh-extra.com/html/biolog-id/smart-storage-agitator-30/smart-storage-agitator-30_user-manual_2753942010.webp)
User Manual Smart Storage Agitator 30
Version
: 1.0
Validation date of the manual: 01/08/2021
10 of 29
The operating principle of the RFID system is based on a transponder (RFID tag) and an
interrogator (coupler). The latter is an active radiofrequency transmitter device that will
activate the RFID tags located in the slot by providing them with the energy they need to
operate. In addition to the energy, the interrogator sends specific commands to which the
RFID tag responds. For example, the interrogator can ask the tag to return the matching
donation number associated to a unique identifier.
2.3.4
Temperature sensor
The temperature sensor integrated in the SST-A30 is waterproof. It allows to measure the
temperature of an area.
Fig. 7: Temperature sensor.
Only the temperature of the climatic chamber (incubator) is the reference. SST-A30 has no
claimed performance regarding temperature.
The SST-A30 gives the temperature only as an indication. This function does not provide the
safety associated with maintaining the storage temperature of platelet concentrates.
2.3.5
Agitator sensor
The Agitation sensor (accelerometer) integrated in the SST-A30 set allows to record the
shaking and stopping phases of the agitator.
2.4
Hardware and software compatibility
2.4.1
Agitator
The SST-A30 is compatible with the Helmer PF48 agitator.
For more information, please contact Biolog-id's Quality Department at
2.4.2
Climate chamber
The SST-A30 is compatible with all environmental chambers used for the storage of the PC
bags if the platelet agitator is not an incubator.