27
7 Troubleshooting
Fault
Code
Cause
Effect
Actions
Information codes
102 INFO_CREDITS_LOW
None
Time to refill
credits
103 INFO_CREDITS_OUT
The BCD can’t
be used
Refill credit
Warning codes
300 WARNING_THERMISTOR_TEMP_HIGH
None.
Treatment is
continued
Cool down BCD
before starting
new treatment
301 WARNING_THERMISTOR_TEMP_LOW
None.
Treatment is
continued
Warm up BCD
before starting
new treatment
302 WARNING_TO_MANY_RED_LEDS_BROKEN
Desired effect
may not occur
Contact Biolight
service centre
303 WARNING_TO_MANY_IR_LEDS_BROKEN
Desired effect
may not occur
Contact Biolight
service centre
Error codes
500 ERROR_NO_SUCH_CODE
Restart BCD
If code reoccurs
contact Biolight
service centre
501 ERROR_SYNTAX_ERROR
Restart BCD
If code reoccurs
contact Biolight
service centre
550 ERROR_TREATMENT_PARAMETER_SECTIONTIME
Treatment
corrupted
Contact Biolight
service centre
551 ERROR_TREATMENT_PARAMETER_FLASHFREQ
Treatment
corrupted
Contact Biolight
service centre
552 ERROR_TREATMENT_PARAMETER_DUTYCYCLE
Treatment
corrupted
Contact Biolight
service centre
553 ERROR_TREATMENT_PARAMETER_HOLDTIME
Treatment
corrupted
Contact Biolight
service centre
554 ERROR_TREATMENT_PARAMETER_NROFSECTIONS
Treatment
corrupted
Contact Biolight
service centre
303181_M17001_ver0_Biolight_Equine_Manual_Animal_ENG_A5.indd 27
2017-12-07 15:17