18
Epic Hygiene
User Manual
Figure 6.3: Wireless Footswitch
WIRELESS FOOTSWITCH
The Epic Hygiene will not emit laser energy until the user
presses down on the Footswitch while the laser is in
Ready
mode. The Footswitch is designed to work using
wireless technology and is paired to the laser console at
the factory.
REMOTE INTERLOCK
This feature allows the laser to be connected to a remote
sensor which prevents it from firing when the sensor is
triggered. To install the remote interlock, insert the plug
(a)
at the end of the connector
(Figure 6.4)
into the rear
of
the laser console and attach the two wires
(b)
at the other
end to a door switch; the laser will stop immediately when
the connection to the door switch is deactivated, i.e.,
when the door is opened.
EMERGENCY STOP
Press the red emergency laser stop button to instantly
turn off the laser console
(Figure 6.5)
. The message
“Emergency Switch Error” will appear
on the console
display and the amber LED will begin flashing. To clear
the error, press the emergency laser stop button again; in
2 to 5 seconds the amber LED will stop flashing and the
system will automatically go into
Standby
mode.
FUNCTIONAL DISPLAY
The system color display with touch screen and LED
indicators on the control panel shows the functional
conditions of the system.
Figure 6.4: Remote
Interlock Connector
Figure 6.5: Emergency
Laser Stop
CAUTION:
Use only the
DC P
ower
S
upply
M
odule supplied with the
Epic Hygiene
laser
system. (BIOLASE P/N 2400
075
)
6
Safety with the
Epic Hygiene
(a)
(b)
Summary of Contents for Epic
Page 2: ...PAGE LEFT BLANK ...
Page 51: ...Appendix F Electromagnetic Compatibility 51 Epic Hygiene User Manual ...
Page 52: ...Appendix F Electromagnetic Compatibility Epic Hygiene User Manual 52 ...
Page 53: ...Appendix F Electromagnetic Compatibility Epic Hygiene User Manual 53 ...
Page 55: ...PAGE LEFT BLANK ...