6.3.3 Primary Tubes and Serum cups
You can also use primary tubes and serum cups in a sam
•
12mm tubes: use the normal tray and put cups directly in the tray
•
16mm tubes: use specific tray, and adapters for cups.
N
N
o
o
t
t
e
e
:
:
The serum level sensor must be enabled, and you must be very careful to pour a volume big enough
for the analyses you want. Otherwise you risk damaging the probe.
6.3.4 Primary tubes with Barcode reader
You can use the barcode reader with primary tubes of 12 or 16. The length of the tube has to be 100mm.
•
Enter the patients with Barcode number as ID.
•
Place the barcode on the tube, turned outwards.
cannot be moved once placed.
•
In the Start Menu, click on
the positions, and affect the analyses
6.4 Prepare Reagents
You can easily prepare the reagent by
To know which reagents are missing, read the flag
Blue Flag: Reagent R1, volume sufficient.
Yellow Flag: Reagent R2, volume sufficient.
Red Flag: Volume insufficient.
Primary Tubes and Serum cups
You can also use primary tubes and serum cups in a same time:
12mm tubes: use the normal tray and put cups directly in the tray
16mm tubes: use specific tray, and adapters for cups.
The serum level sensor must be enabled, and you must be very careful to pour a volume big enough
nt. Otherwise you risk damaging the probe.
Primary tubes with Barcode reader
You can use the barcode reader with primary tubes of 12 or 16. The length of the tube has to be 100mm.
Enter the patients with Barcode number as ID.
Place the barcode on the tube, turned outwards. The tubes can be placed wherever you want, but
cannot be moved once placed.
In the Start Menu, click on
. The instrument will check all the samples barcodes, register
the positions, and affect the analyses asked in the patient menu.
by looking for there position in the main menu:
To know which reagents are missing, read the flags as:
Blue Flag: Reagent R1, volume sufficient.
Yellow Flag: Reagent R2, volume sufficient.
- 42 -
The serum level sensor must be enabled, and you must be very careful to pour a volume big enough
You can use the barcode reader with primary tubes of 12 or 16. The length of the tube has to be 100mm.
The tubes can be placed wherever you want, but
. The instrument will check all the samples barcodes, register
Summary of Contents for KENZA 240 TX
Page 5: ...5 1General Information...
Page 10: ...10 2Installation...
Page 14: ...14 3Preliminary operations...
Page 17: ...17 4Methods programming...
Page 30: ...30 5Calibrations and controls...
Page 37: ...37 6Prepare the run...
Page 47: ...47 7During the run...
Page 51: ...51 8After the run...