Gene Explorer Thermal Cycler Operation Manual Appendix 3
-
26
-
Appendix 3 Product EMC Statement
Class-I Type-A Equipment
If the instrument is used under dry environment, especially under dry environment where artificial
materials (artificial fabrics, carpet, etc.) present, electrostatic discharge may be generated which may
lead to false conclusion.
The instrument is strictly prohibited from using near strong radiation source; otherwise, normal
equipment operation may be interrupted.
EMC should be in line with the requirements of GB/T 18268.1-2010 and GB/T 18268.26-2010.
1.
Requirements for immunity:
Table 1 Requirements for equipment immunity
Port
Test item
EMC basic
standard
Test value
Performance
criterion
Shell
Electrostatic
discharge (ESD)
GB/T 17626.2
Air discharge:2kV,4kV,8kV;
Contact discharge:2kV,4kV
B
Radiation magnetic
field
GB/T 17626.3
3V/m,80MHz-2.0GHz,80%AM
A
Rated PFMF
a
GB/T 17626.8
3A/m,50Hz
A
AC
Voltage sag
d
GB/T 17626.11
1 cycle 0%;
B
5 cycle 40%
C
25 cycle 70%
C
Voltage interruption
d
GB/T 17626.11
5%, duration:250 cycle
C
Pulse group
GB/T 17626.4
1kV(5/50ns.5kHz)
B
Surge
GB/T 17626.5
Line
to
ground:2kV/
Line
to
ground:1kV
B
Radio
frequency
transmission
GB/T 17626.6
3V,150kHz-80MHz,80%AM
A
DC
c
Pulse group
GB/T 17626.4
1kV(5/50ns.5kHz)
Not
applicable
Surge
GB/T 17626.5
Line
to
ground:2kV/
Line
to
ground:1kV
Not
applicable
Radio
frequency
transmission
GB/T 17626.6
3V,150kHz-80MHz,80%AM
Not
applicable
I/O signal
b
Pulse group
GB/T 17626.4
0.5kV(5/50ns.5kHz)
Not
applicable
Surge
GB/T 17626.5
/
Not
applicable
Radio
frequency
transmission
GB/T 17626.6
3V,150kHz-80MHz,80%AM
Not
applicable
I/O
signal
to
main
power
supply
Pulse group
GB/T 17626.4
1kV(5/50ns.5kHz)
Not
applicable
Surge
GB/T 17626.5
/
Not
applicable