CL-10 Plus – User Manual
52 / 75
ECPUS vers 0.2 eng
•
Mixer volume:
to select the mixing chamber volume (where the reaction takes place).
Only one is the correct option:
a)
High
Volume
= 1,2 mL
All methods use this volume setting. All the other options (medium volume, low volume) refer
to previous CL-10 versions.
•
Sample Pump Timing
: This function is not active.
•
Misc. parameters
. This function is not active.
•
Rest parameters:
This function is used to define rest mode parameters (§ 4.6.2)
∗
temperature:
Temp.(°C)
; normally set at 37°C (in any case between 20 - 40 °C)
∗
time interval between consecutive cycles:
Period (min)
; normally set at 90 min
∗
the cycle type used in the rest mode is the same displayed on the program.