CL-10 Plus – User Manual
15 / 75
ECPUS vers 0.2 eng
2.8.
Changing of application within the same working session.
ATTENTION: Steps 2 and 3 of the following procedure allow recovery of not utilized enzyme, still inside the
starter tube (position 6). If the enzyme vial is empty or all the tests are made, it is not necessary to recover the
enzyme; it is therefore recommended to neglect steps 2 - 3 and to go from step 1 to 4.
1.
Replace R1 buffer, reconstituted as required by the method in use, with the R1 buffer,
reconstituted as required by the next method. Leave the enzyme vial (pump 6) of the
method in use in its place.
2. From
File
menu,
Select
the ENZY_REC_R3 method.
3. Run
Sample
once using the GO icon or the F5 function key. Press
Start Measure
and
Accept
. Wait until the end of the run.
4.
Replace the enzyme vial (pump 6) with a vial containing distilled water.
5. Run
Prime Enzyme
twice using the prime enzyme icon or the F2 function key.
6. Run
Clean
once using the clean icon or the F3 function key
.
7.
Wipe the starter needle going to enzyme pump 6 and insert it into the enzyme vial
(starter) of the next chosen method.
8. From
File
menu,
Select
the next method.
9.
Empty the waste bottle.
10.
Follow the instructions given in the package insert of the method to use.